Troubleshooting                            

 

                                 

A12.1095

 

6.1 Optical Part 

TROUBLE

 

CAUSE

 

SOLUTION

 

The edge of the field 
of view is dark or 
the brightness is not 
uniform 

The nosepiece is not in the located 
position( The objective is not in the 
center of the light path ) 

Adjust it into the located 
position( turning the objective to 
let it in the center of the light 
path correctly) 

The condenser is not centered 

Adjust it to center 

A lens (the objective, condenser, 
eyepiece or collector) is dirty. 

Clean it thoroughly 

 
  Find dust and stain 
in the field of view 

A lens (the objective, condenser, 
eyepiece or collector) is dirty. 

Clean it thoroughly 

There are stains on the slide 

Clean it thoroughly 

The position of condenser is too low 

Adjust the height of the 
condenser 

 
 
 
 
 
 
Bad image quality 
(low-resolution, bad 
Contrast) 

No cover glass on the specimen 

Add    cover glass 

The cover glass is too thick or too 
thin 

Use standard cover glass with 
thickness of 0.17mm 

The specimen is on the reverse side 

Turn it over 

Dry objective has oil on it. 
(especially for 40X objectives) 

Wipe the oil 

A lens (the objective, condenser, 
eyepiece or collector) is dirty. 

Clean it 

Immersion oil is not used with the 
100x objective 

Use specified oil 

Air bubbles existed in the 
immersion oil 

Eliminate the bubble 

Use the unspecified oil 

Use the specified oil 

The opening of Aperture diaphragm 
is too large 

turn it down to the proper size 

There are stains on the incidence 
lens of the binocular head 

Clean the lens 

The opening of Aperture diaphragm 
is too small 

Opening it to the proper size 

The position of condenser is too low  Adjust its position 

 
One side of the 
image darkens 

The condenser is not in the center of 
the field or the condenser incline 

Reset the condenser and adjust 
the Central bolt of the condenser 
carefully 

The nosepiece is not in the located 
position 

Rotate the nosepiece to the 
required position 

The specimen is in the floating state 

Fix it firmly 

The image move   
when focus it

 

The specimen is floating on the 
stage

 

Fix it    firmly

 

The nosepiece is not in the located 
position

 

Rotate the nosepiece to the 
required position

 

Содержание A12.1095

Страница 1: ...ed Instruction Manual This manual is for users of A12 1095 scientific research microscope To ensure your safety obtain optimum performance and to familiarize yourself fully with the use of this micros...

Страница 2: ...ing the extensional function board 11 3 11 Computer control signal interface and photo function key 11 3 12 Additional function keys 12 3 13 Touch screen 12 4 Installation 13 4 1 Installing Diagram 13...

Страница 3: ...nvection air which is used for cooling 6 Make sure the instrument is earthed to avoid lightning strike 7 Check the input voltage be sure the input voltage which signed in the back of the microscope is...

Страница 4: ...or potential sources of electrical sparks and use them in a drafty room as possible as you can 5 Don t use organic solvent to wipe the non optical elements when you need to clean use the soft detergen...

Страница 5: ...4 1 Components Name A12 1095 Objective Eyepiece Viewing Head Condenser Nosepiec e Main Body Stage...

Страница 6: ...peed control button 5 Condenser adjustment knob 6 Switch button for upper and lower light 7 Brightness adjustment knob 8 Condenser 1 2 3 4 5 6 7 8 9 10 11 12 13 9 LCD touch screen 10 Aperture iris dia...

Страница 7: ...r condenser 3 Condenser adjustment knob 4 Adjustment hand wheel 5 Photo button 6 Filed iris diaphragm adjustment ring 7 Shortcut 8 Tension adjustment ring for lamp housing 9 Ground lever 10 Lamp housi...

Страница 8: ...t on the top is opening the symbol is When the light switch button is push out the light on the bottom is opening the symbol is 3 3 Light socket Fig 3 As shown in Fig 3 there are red dots in the socke...

Страница 9: ...e the objective may collide with the specimen 3 5 Adjusting the interpupillary distance Fig 5 The interpupillary distance range 47mm 78mm While looking through the eyepieces adjust for binocular visio...

Страница 10: ...e converted to other magnification objective to achieve focusing without damaging the specimen 2 When camera is needed Push the light path switch selector rod as shown in Fig 7 to the end for binocula...

Страница 11: ...he field iris diaphragm inscribed with the field of viewing That means the condenser is centred correctly 7 Slightly increase the field iris diaphragm in reality operation to make the image circumscri...

Страница 12: ...out the standard bright and dark field board then inset the extensional function board into the slot and tighten the screw 3 11 Computer control signal interface and photo function key Fig 12 13 Conn...

Страница 13: ...croscopes has a set of supporting software For the specific operation of the software please refer to the software manual The function keys can be used in combination with the touch screen Please refe...

Страница 14: ...installation sequence of the components The number in the figure shows the assembly steps Before installing be sure every components is clean do not score any parts or glass surface Keep well with hex...

Страница 15: ...l groove of the stage holder with the dovetail bulge of the frame and slide down naturally until lower limit screw is limited Then the screw I in Fig 2 is screwed inward with a wrench 4 2 2 Installing...

Страница 16: ...photo attachment or C mount 0 5 Mount the photo attachment or 0 5 C mount according to the assembly path into the connecting seat of the trinocular viewing head and tighten the screw V During the inst...

Страница 17: ...en instruction manual for specific operation Clean the objectives regularly for lens is susceptible to dust When operating use 10 magnification objective to search and focus specimen firstly then repl...

Страница 18: ...erface into the ternary camera barrel to an appropriate position Then screw in the number XI bolt to fasten the camera interface In order to ensure the appropriate position of camera interface the spe...

Страница 19: ...ctric stage 270 210mm Moving range 125 75mm Illumination 3W LED lamp Condenser Abbe condenser NA0 9 0 1 Software Image software Scopeimage10 Photo attachment Video attachment C Mount 1 C Mount 0 5 Pol...

Страница 20: ...the reverse side Turn it over Dry objective has oil on it especially for 40X objectives Wipe the oil A lens the objective condenser eyepiece or collector is dirty Clean it Immersion oil is not used wi...

Страница 21: ...en display magnification The objective is not properly installed to the specified converter hole or the screen setting is inconsistent with the actual situation The objective is re installed to the co...

Страница 22: ...the instructions A12 1095 Stage is not working properly Improper line connection Check the internal connections and reconnect well Initialization cannot stop after turning on Improper control board R...

Отзывы: