29
ELECTROMAGNETIC COMPATIBILITY INFORMATION
Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration - electromagnetic emissions
The HZ-8595 is intended for use in the electromagnetic environment specified below.
The customer or the user of the HZ-8595 should assure that it is used in such an environment
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The HZ-8595 uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11
Class B
The HZ-8595 is suitable for use in all establishments other than domestic
and those directly connected to the public low-voltage power supply net-
work that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions IEC 61000-3-3
Not applicable
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration - electromagnetic immunity
The HZ-8595 is intended for use in the electromagnetic environment specified below. The customer or the user of the HZ-8595 should assure that
it is used in such an environment.
IMMUNITY test
IEC 60601test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical commer-
cial or hospital environment.