TROUBLESHOOTING
Power supply
Battery powered mode:
6VDC 4*AAA batteries
AC adaptor powered mode: 5V 1A (Not included)
(Please only use the recommended AC
adaptor model).
Display mode
Measurement mode
Oscillographic testing mode
Measurement range
Accuracy
Normal working condition
Storage & transportation
condition
Measurement perimeter
of the upper arm
About 22cm~32cm
Net Weight
External dimensions
Attachment
4*AAA batteries,user manual,
Mode of operation
Continuous operation
Degree of protection
Protection against
ingress of water
Software Version
A01
Pressure:
5°C-40°C within±3mmHg(0.4kPa)
Pulse value:±5%
Device Classification
Battery Powered Mode:
Internally Powered ME Equipment
AC Adaptor Powered Mode:
Class II ME Equipment
Rated cuff pressure:
0mmHg~299mmHg(0kPa ~ 39.9kPa)
Measurement pressure:
SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa)
DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)
Pulse value: (40-199)beat/minute
IP21 It means the device could protected against
solid foreign objects of 12.5mm and greater, and
protect against vertically falling water drops.
Type BF applied part
WARNING: No modification of this equipment is allowed.
Digital LCD display V.A.60mm*41mm
Approx.162g(Excluding the batteries)
Approx.93mm*130mm*32.5mm
A temperature range of :+5°C to +40°C
A relative humidity range of 15% to 90%,
non-condensing, but not requiring a water
vapour partial pressure greater than 50 hPa
An atmospheric pressure range of :
700 hPa to 1060 hPa
Temperature:-20°C to +60°C
A relative humidity range of ≤ 93%, non-condensing,
at a water vapour pressure up to 50hPa
SPECIFICATIONS
This section includes a list of error messages and frequently
asked questions for problems you may encounter with your blood
pressure monitor. If the products not operating as you think it
should, check here before arranging for servicing.
PROBLEM SYMPTOM CHECK THIS
REMEDY
No power
Low
batteries
Error
message
Display will not
light up.
Batteries are exhausted. Replace with new batteries
Insert the batteries
correctly
Replace with new batteries
Batteries are inserted
incorrectly.
Display is dim or
show
Batteries are low.
EExx,shows on
the display.
A calibration error
occurred.
Retake the measurement.
If the problem persists,
contact the retailer or our
customer service
department for further
assistance.Refer to the
warranty for contact
information and return
instructions.
AC adaptor is inserted
incorrectly.
Insert the AC adaptor
tightly
Warning
message
“out ” shows
Out of measurement
range
E 01 shows
E 03 shows
E 04 shows
The treatment of the
measurement failed.
E 02 shows
The cuff is too tight
or too loose.
Refasten the cuff and then
measure again.
The monitor detected
motion, talking,or the
pulse is too poor
while measuring.
Relax for a moment and
then measure again.
Relax for a moment and
then measure again.
The measurement
process does not
detect the pulse
signal.
Loosen the clothing on
the arm and then
measure again.
Relax for a moment.
Refasten the cuff and then
measure again. If the
problem persists, contact
your physician.