13
CONNECTORS
Connecting load cells to the indicator
● The indicator can connect with 6 load cells of 350Ω at most
● 4 wire or 6 wire load cell connections are both okay
● Please contact us directly if you have other special needs for your application
● There are two connection methods between the load cell and indicator
Quick Disconnect as shown below: Figure 2: Quick Disconnect connection diagram
Figure 2:
Quick Disconnect
connection diagram
Hardwire (Using Inner Terminal Block Connection:
Note: Make sure you follow all the anti-static rules to avoid damage to your indicator
● Excitation voltage: 5V DC
● Largest output current: 120 mA
● Excitation circuit: 5 VDC, 4 wire connection, 6 load cell of 350ohm maximum
● Open the back cover of the weighing indicator, and insert signal cable to the
terminal block (see figure 3); Make sure the screw on terminal block is fixed tightly
FIGURE 3:
INNER TERMINAL BLOCK
CONNECTION DIAGRAM
Table 4. Wiring Color Code
Signal Name
Color Code
Description
+Exe/ +EX
RED
Positive excitation voltage to load cell
+IN / +SIG
GREEN
Positive output signal from load cell
HD / SHLD
YELLOW/THICK BLACK
Shield Wire
-IN / -SIG
WHITE
Negative output signal from load cell
-EXC / -EX
BLACK
Negative excitation voltage to load cell