![Optima Scale OP-900-LD Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/optima-scale/op-900-ld/op-900-ld_user-manual_1642670010.webp)
8
20. Press ENTER
to continue. The display will now show
C06
.
21. Press ENTER
to access the C06 channel. The display will show [
C6 0
].
22. The C6 channel is used to calibrate the scale with a known weight. Press ZERO
to set the value of C6 to [
C6 1
]. Press ENTER
. The display will flash
SPAN
,
and then show [######].
23. Enter the calibration weight value you will use (at least 10% of max capacity you set
in C04 by using TOTAL and UNIT to move the cursor left and right, and TARE
and ZERO move the values down and up.
24. Place the calibration weight you have on the empty scale and press ENTER
.
25. The scale will count down from 10 to 0. Once 0 has been reached, the display will
show
CALEnd
.
26. Press ENTER
to continue. The display will now show
C07
.
27. Press ON/OFF
quicly
to save and exit the setup menu.
28. The scale has now been calibrated. The display will show the value of the calibration
weight on the scale.
29. If the scale does not show the value of the calibration weight, check that the feet on
the platform are not screwed in too tightly, and verify that the platform is level.
30. Unload the scale; the display should read
000000
.
31. If the scale does not display 000000, check that the feet on the platform are not
screwed in too tightly, and verify that the platform is level.
REMOTE CONTROL
Button
Sign
Function
1
ON/OFF
2
Total
3
Unit
4
Tare
5
Zero
6
Enter/Set