4.    Controls and Functions 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Joystick Left/Right:

  Selects 

Zoom

Brightness

 and 

Focus

 functions. 

 

Joystick Up/Down:

  Increase/Decrease value of the selected function. 

 
Auto B:  

Switches between auto brightness and manual brightness modes.                 

 

-  When 

Auto B

 is activated, brightness of the picture is controlled automatically. Joystick 

up/down will change “target” value of the brightness adjustment control.   

-  When 

Auto B

 is turned off, the camera goes to manual brightness control. Level of the 

image brightness can be adjusted manually by Joystick up/down. 

 
Auto F: 

 

Switches between automatic focus and manual focus control modes. 

 

-  When 

Auto F

 is activated, camera will adjust the focus automatically. 

-  When 

Auto F

 is turned off, image focus is adjusted manually by Joystick up/down. 

 
Note:

 Image brightness needs to be set to sufficient level for smooth 

Auto F 

function. 

Note:

 Working distance of the camera shall be set right. Camera cannot focus on the object 

if working distance is out of the specified range.  

 
Iris 

mode

:

 

 

Push this button shortly to activate 

Iris mode

 

 

When 

Iris mode

 is activated, camera will fix the Gain and Exposure time and allows 

user to control Iris (aperture) of the camera. Iris size (aperture diameter) is adjusted by 
Joystick up/down. Make sure 

Brightness

 function is selected to adjust Iris.  

 

Note:

  Iris adjustment allows the user to control depth of field (dept of focus). Decreasing Iris 

size will produce images with larger depth of field.  

Note:

  Image quality is improved in most cases when Iris size is decrease due to less 

aberration.  

Note:

 Image brightness is lowered, why more illumination is needed when Iris is decreased.  

 

Содержание OP-009 020

Страница 1: ...Instructions Manual for Control Unit for Optilia High Definition Cameras Product No OP 009 020 This manual describes how to use Control Unit together with W10x HD W20x HD HD10 and HD20 cameras ...

Страница 2: ...tes with Optilia HD cameras via serial RS232 communication protocol 2 Connections Power On Sequence a Attach circular connector of HD camera to the mating connector on Control Unit b Attach the Foot Switch to external foot switch input of the Control Unit Optional c Connect the Power supply to 12 5 V DC input of Control Unit d Switch On the Control Box After switching on the Control Box it will lo...

Страница 3: ... focus is adjusted manually by Joystick up down Note Image brightness needs to be set to sufficient level for smooth Auto F function Note Working distance of the camera shall be set right Camera cannot focus on the object if working distance is out of the specified range Iris mode Push this button shortly to activate Iris mode When Iris mode is activated camera will fix the Gain and Exposure time ...

Страница 4: ...our are recorded in camera Press R shortly to recall recorded adjustments any time All rights reserved The information contained herein is designed only for use with Optilia Instruments AB HD Inspection Cameras Optilia Instruments AB is not responsible for any use of this information for other purposes Optilia Instruments AB shall not be liable to the purchaser of this product or third parties for...

Отзывы: