Page 10
F
Fig. 11
ig. 11
F
Fig. 13
ig. 13
FFig. 12
ig. 12
①
8.
Remove the mirror, place the specimen on the stage and
adjust the focus.
9. Adjust the background contrast by turning the prism movement
knob
②
on the DIC slider. (Fig. 11)
10.
When the prism movement knob
②
on the DIC slider is turned,
the interference color in the background varies continuously
from the gray to magenta (from -100 to 600 nm). Set the
interference color which can provide best contrast with respect
to the specimen.
•
If the background color is dark, the observation is similar to
darkfield.
•
If the background color is grey, three-dimensional high contrast
images can be achieved rich in grey shades.
•
If the background color is close to magenta, even small phase
variation can be observed as a change in color.
②
9.3 IM-5MET
• As with observation in polarized light, DIC observation
cannot be performed when the BF/DF switch on the mi-
croscope is in the ‘DF’ position.
1.
Move the lever in “BF” position for brightfield illumination.
2.
Insert the polarizer “PO” with the inscription facing the user.
(Fig. 12)
3.
Move the polarizer “PO” into the light path until the slider
clicks into its position.
4.
Insert analyzer “AN” with the inscription facing up. (Fig. 13)
5.
Move the rotatable analyzer “AN” into the light path until the
slider clicks into its position.
6.
Put a flat mirror on the stage and get into approximate focus
the mirror surface.
7.
While observing into the eyepieces, rotate the scale of the
analyzer
①
until you achieve the darkest image. (Fig. 14)
•
Now the light extinction (so-called “crossed Nicol position”)
is achieved.
8.
Remove the mirror, put a specimen on the stage and focus
the specimen.
9. Adjust the background contrast by turning the prism movement
knob on the DIC slider
②
. (Fig. 11)
10.
When the prism movement knob
②
on the DIC slider is turned,
the interference color in the background varies continuously
from the gray sensitive color to magenta sensitive color (from
-100 to 600 nm). Set the interference color which can provide
best contrast with respect to the specimen.
•
If the background color is dark, the observation is similar to
darkfield.
•
If the background color is grey, three-dimensional high contrast
images can be achieved rich in grey shades.
•
If the background color is close to magenta, even small phase
variation can be observed as a change in color.
F
Fig. 14
ig. 14
①
Содержание ACCESSORIES Series
Страница 1: ...Ver 2 1 2022 Model M 870 ACCESSORIES Series INSTRUCTION MANUAL...
Страница 4: ...Page 4 3 Package content DIC Prism...
Страница 12: ...Ver 2 1 2022 Modello M 870 Serie ACCESSORI MANUALE DI ISTRUZIONI...
Страница 15: ...Pagina 15 3 Contenuto della confezione Prisma DIC...
Страница 23: ...Modelo M 870 Serie ACCESORIOS MANUAL DE INSTRUCCIONES Ver 2 1 2022...
Страница 26: ...P gina 26 3 Contenido del paquete Prisma DIC...
Страница 34: ...Mod le M 870 S rie ACCESSOIRES MANUEL D UTILISATION Ver 2 1 2022...
Страница 37: ...Page 37 3 Contenu de l emballage Prisme DIC...
Страница 45: ...Modell M 870 ZUBEH R Serie BEDIENUNGSANLEITUNG Ver 2 1 2022...
Страница 48: ...Seite 48 3 Verpackungsinhalt DIC Prisma...
Страница 56: ...Modelo M 870 S rie ACESS RIOS MANUAL DE INSTRU ES Ver 2 1 2022...
Страница 59: ...P gina 59 3 Conte do da embalagem Prisma DIC...