6.2 Data storage number setting and recalling of stored data
6.2.1 Data storage number setting
For setting a data storage number for the next data storage, take the following procedure.
• Press
MEM
key in the “hold” mode in the standard measurement mode or continuous
measurement mode to move to the data storage mode. The " " mark under the main marker
"
MEM
" will light and "
NO
" will be displayed in the sub marker. Also, the data storage number
"
1
" will be displayed in the sub display for the initial data storage. When measured data have been
already stored in any data storage number, the "
next number
" to the last data storage number will
be displayed.
• By pressing
△
or
▽
key in the hold mode, the
least significant digit will blink for setting it.
• Set the desired figure by
△
or
▽
key and press
ENT
key for shifting the setting digit to the next
higher digit.
•
Repeat the above procedure up to the most
significant digit.
• By pressing
ENT
key at the most significant digit,
the setting of data storage number will be
completed.
6. Temperature data storage.
*
Make sure to press any keys until you hear an audible beep.
Tb CONT MEM PEAK
MEAS
HOLD
AL
AH
C
F
MAX
AVE
ε
r
MIN
TC
NO
Main
marker
Main
display
Sub
display
Sub
marker
• When any measured data has not been stored in the setting data storage number,
"
non
" will be displayed in the main display.
Reference
• When a measured data has been stored in the setting data storage number, its
temperature data will be displayed in the main display.
Reference
• On condition that a data storage number has been set by the above procedure,
when you press
MEAS
key to start a measurement and then press
ENT
key, the
measured data already stored in its data storage number will be overwritten by a
new measured data.
• If you do not want to overwrite the measured data already stored in its data
storage number, follow the above procedure again to select the data storage
number that any measured data has not been stored in.
Caution