5.6.2 High temperature alarm setting     

This setting is for a judgment of high alarm during measurement. When the high alarm is judged, the 

status marker "

AL

" will light and the buzzer will sound. When the "

oFF

" is selected, neither the alarm 

judgment can be executed nor the buzzer can be sound. 

• Press 

MEAS

 key for about 1 second to turn on the power supply and then release 

MEAS

 key. The 

status marker "

HOLD

" will light. Then press 

SEL

 key for about 2 seconds in the hold mode to move to 

the setting mode. A setting value will be displayed in the main display and its mode will be displayed in 

the sub display. Press 

SEL

 key to display "

AH

" in the sub display. By pressing 

 key, the least 

significant digit will blink to set it. 

• Set the desired figure by 

 or 

 key and press 

ENT

 key for shifting the setting digit to the next higher 

digit. 

• Repeat the above procedure up to the most significant 

digit. 

• By pressing 

ENT

 key at the most significant digit, the 

setting value will be stored. 

 

 
 
 

 
 

                                             

 

 

 

5.6.3 Modulation value setting     

When the "

dELy

" is selected in the signal modulation mode, the first-order lag degree can be adjusted by 

setting of the modulation time constant. 

When the "

PEAk

" is selected, the damping degree of signal after tracing the peak value can be adjusted.   

• Press 

MEAS

 key for about 1 second to turn on the power supply and then release 

MEAS

 key. The 

status marker "

HOLD

" will light. Then press 

SEL

 key for about 2 seconds in the hold mode to move to 

the setting mode. A setting value will be displayed in the main display and its mode will be displayed in 

the sub display. 

5.6.3-1) Modulation time constant selection 

(effective when the "dLEy" is selected in the signal

 

modulation mode) 

• Press 

SEL

 key to display "

tAu

" in the sub display. 

• By  pressing 

 or 

 key, the modulation time 

constant will blink in order of 0.0 

 0.2 

 0.5 

 1.0 

(second) in the main display. Select the desired 
modulation time constant and press 

ENT

 key. 

 

 

 

 

             

5.7.3-2) Damping degree selection

 

(effective when the "PEAk" is selected in the signal modulation 

mode)

 

• Press 

SEL

 key to display "

dEc

" in the sub display.                             

• By pressing 

 or 

 key, the modulation time constant 

will blink in order of 0 

 2 

 5 

 10 (ºC/second) 

(ºF/second) in the main display. Select the desired damping 

degree and press 

ENT

 key. 

 
 
 

 

5. Measuring

                                 

*

Make sure to press any keys until you hear an audible beep.

 

   Tb  CONT  MEM  PEAK

 

 

MEAS 

 

HOLD

  AL   

 

AH

C

F

 MAX

 

AVE

 

ε

r

 MIN

  

TC

  

NO

 

Main display

 

Sub display 

Main display

 

   Tb   CONT  MEM  PEAK

 MEAS 

HOLD

  AL 

 

AH

C

F

 MAX

 

AVE

 

r

 MIN

  

TC

  

NO

 

Sub display

 

Main display

 

Sub display

 

   Tb   CONT  MEM  PEAK

 MEAS 

HOLD

  AL 

 

AH

C

F

 MAX

 

AVE

 

r

 MIN

  

TC

  

NO

• The setting range is “oFF” or 400 to 3000ºC/752 to 5432ºF. The initial is "

oFF

". 

Reference

[How to reset the low temperature alarm to "oFF"]

• Press 

SEL

 key for about 2 seconds to display "

AH

" on the sub display. By changing the 

displayed numeric figure to be lower than 399ºC (751ºF) by 

 , 

 and 

ENT

 keys,

"

oFF

" will be displayed.  Press 

ENT

 key to store it. 

Reference

• If 

SEL

key is pressed for 2 seconds in the setting mode or if any key is not pressed for 

1 minute, the thermometer will return to the measurement mode. 

R e m a r k s

• The initial is "

0.0

"

 

second. (The displayed value is based on the real signal without any 

modulation.) 

R e m a r k s

• If 

SEL

key is pressed for 2 seconds in the setting mode or if any key is not pressed for 

1 minute, the thermometer will return to the measurement mode. 

R e m a r k s

• The initial is "

0.0

"

 

second. (The displayed value is based on the real signal without any 

modulation.) 

R e m a r k s

• If 

SEL

key is pressed for 2 seconds in the setting mode or if any key is not pressed for 

1 minute, the thermometer will return to the measurement mode. 

R e m a r k s

Содержание THERMO-HUNTER VF-3000

Страница 1: ...e part of Optex 2 No part of this manual may be reproduced or transmitted in any form or by any means electronic or mechanical including photocopying and recording for any purpose other than the purch...

Страница 2: ...econds in the hold mode the power supply is automatically turned off Avoid measuring shiny objects Shiny objects reflect surrounding temperatures The emissivity rate of the thermometer can be adjusted...

Страница 3: ...heating of the batteries may cause fire or injury Never use the thermometer if it has been broken smoking or nasty smelling These may cause fire When the thermometer is broken smoking or nasty smelli...

Страница 4: ...alarm setting 10 5 6 2 High temperature alarm setting 11 5 6 3 Modulation value setting 11 5 6 3 1 Modulation time constant setting for dLEy selected 11 5 6 3 2 Damping degree setting for PEAk select...

Страница 5: ...ect viewfinder enables to measure small objects with distance With the digital display in the viewfinder you can see a measured value while sighting to an object 2 1 Model VF 3000 2 color type 600 to...

Страница 6: ...measured value and parameters are displayed 3 Battery cover Remove it by lightly pushing both sides of the triangle mark and sliding it to the arrow direction Ref 4 1 Loading batteries 4 Lens cap For...

Страница 7: ...NT The mark under CONT will light in the continuous measurement mode CONT MEM The mark under the MEM will light in the data storage mode MEM PEAK When the PEAK is selected in the signal modulation mod...

Страница 8: ...sliding it upward while pushing both sides lightly SUM 3 SIZE 1 5V 2 Battery life The battery warning maker will blink when the battery life becomes shortened Replace the batteries Alkaline AA or UM 3...

Страница 9: ...distance is about 3m The measurement diameter and the inner side of the targeting mark are almost same 3 For the measurement distance is longer than 3m Measurement diameter becomes smaller than the ta...

Страница 10: ...he measured value will be held The status marker MEAS will go off and HOLD will light 5 2 Continuous measurement mode 5 2 1 Start of continuous measurement This mode is for continuous measurement by f...

Страница 11: ...er than the measuring range Temperature displaying oFL 2 color 2020 C 3668 F Single color 3020 C 5468 F Temperature displaying uFL 2 color 580 C 1076 F Single color 380 C 716 F Remarks When you use th...

Страница 12: ...g unit selection Unit C C F F C C 5 5 2 2 color single color selection CoLr 2 1 2 5 5 3 Press MEAS key for about 1 second to turn on the power supply and then release MEAS key The status marker HOLD w...

Страница 13: ...ree setting 5 5 2 Temperature unit selection Select C or F for measuring temperature unit Press MEAS key for about 1 second to turn on the power supply and then release MEAS key The status marker HOLD...

Страница 14: ...a judgment of low alarm during measurement When the low alarm is judged the status marker AL will light and the buzzer will sound When the oFF is selected neither the alarm judgment can be executed no...

Страница 15: ...Press SEL key to display tAu in the sub display By pressing or key the modulation time constant will blink in order of 0 0 0 2 0 5 1 0 second in the main display Select the desired modulation time con...

Страница 16: ...ay Sub display Tb CONT MEM PEAK MEAS HOLD AL AH C F MAX AVE r MIN TC NO Initial data storage Tb CONT MEM PEAK MEAS HOLD AL AH C F MAX AVE r MIN TC NO Main marker Main display Sub display In case of da...

Страница 17: ...ent mode press MEAS key again to move to the hold mode Referring to 5 4 Emissivity rate setting set a new emissivity rate by pressing and ENT keys in the hold mode The new emissivity rate set will be...

Страница 18: ...above procedure up to the most significant digit By pressing ENT key at the most significant digit the setting of data storage number will be completed 6 Temperature data storage Make sure to press an...

Страница 19: ...de Reference When you recall a stored data be careful on the following point Example 1 If the measured data have been stored in the data storage number up to No 50 51 that is the next number to the la...

Страница 20: ...no will stop and no will light To quit this mode press SEL key for about 2 seconds 6 Temperature data storage Make sure to press any keys until you hear an audible beep Tb CONT MEM PEAK MEAS HOLD AL...

Страница 21: ...the working temperature Er1 display only Use the thermometer in the environment from 0 to 50 C 32 to 122o F Main display Thermometer 2 EEPROM data broken Stored data temperature data and calibration d...

Страница 22: ...in the standard measurement mode or continuous measurement mode the mode will become the measurement mode settings mode In the measurement mode settings mode by pressing SEL key for about 2 seconds o...

Страница 23: ...2 F Response Time 0 2 second Emissivity Compensation 1 900 to 0 100 Mathematics Maximum value minimum value average value Signal Modulation Peak delay Display System LCD digital 4 digits Displayed in...

Страница 24: ...E 459 0P2CE Oct 08 Thermo hunter Model VF 3000 Printed in Japan Ver 1 11 03 04First OPTEX CO LTD 5 8 12 Ogoto Otsu 520 Japan TEL 077 579 8680 FAX 077 579 8199 http www optex co jp meas meas optex co j...

Отзывы: