Operation

Normal Measurement

[Starting Normal Measurement]

Press  MEASURE  button to turn on the power supply.  A laser
beam is emitted and the measurement starts.  (In the Normal 
Measurement mode, the unit takes measurements while  
 MEASURE  button is pressed.)
Point the laser beam at a measuring object and aim it at the
center of the area to be measured.  For the distance from this
thermometer to the measuring object, refer to page 9.
In the Normal Measurement mode, the measurement indication
"   " is shown on the LCD display and a laser indicator blinks.

[Quitting Normal Measurement / HOLD mode]

Release  MEASURE  button.  Then, the laser beam turns off and
a value measured last is displayed for 15 seconds (HOLD mode).
The power supply turns off automatically after 15 seconds.

Continuous Measurement

[Starting Continuous Measurement]

Press and hold both of  MEASURE  button and  SET/ENTER  button simultaneously for two seconds or
more, then the Continuous Measurement mode switched on.  (In the Continuous Measurement mode, the
thermometer keeps measuring temperatures without pressing  MEASURE  button.)
During the Continuous Measurement mode, LCD display shows
"

".

[Quitting Continuous Measurement]

To quit the Continuous Measurement mode, press and hold  SET/ENTER  button for two seconds or
more.  Then, it switches to HOLD mode.  The power supply turns off automatically after 15 seconds
in the HOLD mode.

Set the battery supplied as an accessory and operate the unit according to the following procedure:

8

Содержание PT-7LD

Страница 1: ...HUNTER PT 7LD OPTEX FA CO LTD 91 Chudoji Awata cho Shimogyo ku Kyoto 600 8815 JAPAN TEL 81 75 325 1314 FAX 81 75 325 2936 Printed in JAPAN 0806980 2018 6 Portable Non Contact Thermometer Instruction...

Страница 2: ...of View Memory Function Settings Setting High Limit Temperature for Alarm Setting Low Limit Temperature for Alarm Setting Emissivity Ratio HOT COLD mode Setting C F mode Record Erasing Troubleshooting...

Страница 3: ...of controls or adjustments or perfor mance of procedures other than those specified herein may result in hazardous radiation exposure Safe Usage This instruction manual contains various warnings for...

Страница 4: ...things with sharp points which may damage the fil ter Damaged filter causes incor rect measurement KEEP THE THERMOMETER AWAY FROM CHARGED OB JECTS This may cause irreparable dam age or incorrect measu...

Страница 5: ...supply Battery life 0 to 50 Ambient temperature 35 to 85 Rh Without due condensation Ambient humidity 10 to 60 Storage temperature IP67 Protective structure ABS Antibacterial Material H W D 160 44 42...

Страница 6: ...e is no button to turn ON OFF the power supply Pressing MEASURE button wakes up the thermometer The power supply turns off automatically after 15 seconds absence of pressing any button Name of Compone...

Страница 7: ...7...

Страница 8: ...and a value measured last is displayed for 15 seconds HOLD mode The power supply turns off automatically after 15 seconds Continuous Measurement Starting Continuous Measurement Press and hold both of...

Страница 9: ...e away from the measuring object unless there is any obstacle However please note that the measuring field of view enlarges in proportion to the measuring distance This thermometer has an optical reso...

Страница 10: ...g CALL Mode Press and hold CALL button for two seconds or more in HOLD Mode refer to page 8 and CALL sign appears on the LCD display In CALL recalling mode the last record of tem perature data and the...

Страница 11: ...Ratio HOT COLD mode C F mode and Record Erasing settings refer to page 14 When each setting is completed it automatically turns to the next setting mode For example when setting of High Limit Temperat...

Страница 12: ...h LimitTemperature for Alarm cannot be set at lower temperature than Low LimitTempera ture for Alarm Turning ON OFF Alarm Function When the High Limit Temperature setting is completed the setting mode...

Страница 13: ...isplayed value as the Low Limit Temperature for Alarm Attention The Low Limit Temperature for Alarm cannot be set at higher temperature than High LimitTempera ture for Alarm Turning ON OFF Alarm Funct...

Страница 14: ...the Record Erasing press button to select ESC sign on the display and press SET ENTER button Then you can return to the setting mode again Please note that OPTEX is not liable for any loss of data Emi...

Страница 15: ...w alkaline one A sign b blinks Otherwise install the battery correctly in the battery compartment A filter unit is stained Clean the filter referring Filter of MAINTENANCE described in page 16 Field o...

Страница 16: ...for cleaning as it may deform or damage the thermometer Annual Inspection Re calibration is recommended once a year Ask the distributor where you purchased the unit for details Battery Installing Batt...

Страница 17: ...HUNTER PT 5LD OPTEX FA CO LTD 91 Chudoji Awata cho Shimogyo ku Kyoto 600 8815 JAPAN TEL 81 75 325 1314 FAX 81 75 325 2936 Printed in JAPAN 0806970 2018 6 Portable Non Contact Thermometer Instruction...

Страница 18: ...unction Settings Setting High Limit Temperature for Alarm Setting Low Limit Temperature for Alarm Setting Emissivity Ratio DARK BRIGHT mode Setting C F mode Record Erasing Setting Hold mode Selections...

Страница 19: ...of controls or adjustments or perfor mance of procedures other than those specified herein may result in hazardous radiation exposure Safe Usage This instruction manual contains various warnings for...

Страница 20: ...e before taking measurement AVOID MEASURING SHINY OB JECTS DO NOT USE ANY BATTERY OTHER THAN SPECIFIED This may cause irreparable dam age or incorrect measurement DO NOT TOUCH THE FILTER Do not touch...

Страница 21: ...pply Battery life 0 to 50 Ambient temperature 35 to 85 Rh Without due condensation Ambient humidity 10 to 60 Storage temperature IP67 Protective structure ABS Antibacterial Material H W D 160 44 42mm...

Страница 22: ...e is no button to turn ON OFF the power supply Pressing MEASURE button wakes up the thermometer The power supply turns off automatically after 15 seconds absence of pressing any button Name of Compone...

Страница 23: ...7...

Страница 24: ...inuous Measurement Press and hold both of MEASURE button and SET ENTER button simultaneously for two seconds or more then the Continuous Measurement mode switched on In the Continuous Measurement mode...

Страница 25: ...more away from the measuring object unless there is any obstacle However please note that the measuring field of view enlarges in proportion to the measuring distance This thermometer has an optical r...

Страница 26: ...g CALL Mode Press and hold CALL button for two seconds or more in HOLD Mode refer to page 8 and CALL sign appears on the LCD display In CALL recalling mode the last record of tem perature data and the...

Страница 27: ...ttings refer to page 14 HOLD mode Selections refer to page 15 When each setting is completed it automatically turns to the next setting mode For example when setting of High Limit Temperature for Alar...

Страница 28: ...h LimitTemperature for Alarm cannot be set at lower temperature than Low LimitTempera ture for Alarm Turning ON OFF Alarm Function When the High Limit Temperature setting is completed the setting mode...

Страница 29: ...displayed value as the Low Limit Temperature for Alarm Attention The Low Limit Temperature for Alarm cannot be set at higher temperature than High LimitTempera ture for Alarm Turning ON OFF Alarm Func...

Страница 30: ...cel the Record Erasing press button to select ESC sign on the display and press SET ENTER button Then you can return to the setting mode again Please note that OPTEX is not liable for any loss of data...

Страница 31: ...of View described in page 9 The measuring object is smaller than the field of view Adjust the measuring range referring to Field ofView shown in page 9 Affected by a nearby heating source Block the he...

Страница 32: ...for cleaning as it may deform or damage the thermometer Annual Inspection Re calibration is recommended once a year Ask the distributor where you purchased the unit for details Battery Installing Batt...

Страница 33: ...ilter please remove the dust on the lens with a blower etc for lens cleaning use The main unit is made of ABS When it becomes dirty lightly wipe with a soft cloth which has been soaked in soapy neutra...

Страница 34: ......

Страница 35: ...UNTER PT S80 U80 Portable Non Contact Thermometer Instruction Manual 91 Chudoji Awata cho Shimogyo ku Kyoto 600 8815 JAPAN TEL 81 75 325 1314 FAX 81 75 325 2936 Printed in JAPAN 0807000 2018 5 OPTEX F...

Страница 36: ...ction Conformance to EU Directives This in a class A product In residential areas it may cause radio interference in which case the user may be required to take adequate measures to reduce interferenc...

Страница 37: ...se of controls or adjustments or perfor mance of procedures other than those specified herein may result in hazardous radiation exposure Safe Usage This instruction manual contains various warnings fo...

Страница 38: ...Warning Caution KEEP THE THERMOMETER AWAY FROM SUDDEN CHANGE IN AMBIENT TEMPERATURE Sudden temperature change may cause incorrect measurement Start measurement when temperature has become stable after...

Страница 39: ...ED Buzzer ON OFF Selectable Emissivity Adjustment 0 95 0 85 0 70 Selectable Adjustable 0 30 to 1 20 0 01step Temperature unit C F Selectable NOR MAX MIN Display function AA Battery x2pcs Power supply...

Страница 40: ...6 There is no button to turn ON OFF the supper supply Name of Components...

Страница 41: ...7...

Страница 42: ...es supplied as an accessory and operate the unit according to the following procedure 1 Point the laser beam at a measurement objectandaimitatthecenteroftheareato be measured For the distance from thi...

Страница 43: ...Selection of display function Default value for each setting at the time of shipment PT S80 PT U80 Temperature unit C NOR Display function Maximum temperature alarm Temperature setting 600 C Function...

Страница 44: ...eisanyobstacle However pleasenotethatthemeasuringfieldofview depends on measuring distance This thermometer has an optical resolution of 33 1 D Distance to the measuring object S Spot size For Correct...

Страница 45: ...g that AL H on the sub display is changed to AL L in step 2 and later When the lower limit setting has been completed the mode is changed to HOLD When the upper and lower limit alarms are set HI and L...

Страница 46: ...12 PT S80 P 13 Recording Measured Temperature Record Measured Temperature MEM Mode Call Temperature Record CALL Mode P 13 Emissivity Setting...

Страница 47: ...bjects has their own emissivity which changes depending on the surface conditions and object temperature This thermometer has 3 fixed emissivities Refer to the following examples 0 95 Food rubber plas...

Страница 48: ...ime Setting P 17 Emissivity Setting P 18 Recording Measured Temperature Record Measured Temperature MEM Mode Call Temperature Record CALL Mode Delete All the Temperature Records P 19 Continuous Measur...

Страница 49: ...SET button is pressed Display TIME on the LCD After TIME blinks YEAR blinks Then the display is automatically shifted to the state that YEAR lights up and the dominical year blinks After the set year...

Страница 50: ...ght 2 digits minute light up Press the MAX MIN or MEM CAL button to adjust the hour Then press the MEASURE button to fix the setting After the setting the right 2 digits of the clock start to blink fo...

Страница 51: ...ect other than the measured object such as a high temperature object exists on the periphery temperature for the other object is reflected which will cause incorrect measurement Thus it is necessary t...

Страница 52: ...capacity FULL lights up on LCD 1 Press this button for 2 seconds or more while the power is ON Call temperature record CALL mode CALL is displayed on the sub display and then the latest stored temper...

Страница 53: ...ng in page 7 CONT is lit on LCD and continuous measurement is started Measurement continues even if releasing the MEASURE button 1 When the MEASURE button is pressed again or the DIP switch is changed...

Страница 54: ...ing of data inside thermometer 2 PC can read all date instantly on thermometers continuous mode 3 Output all data by CSV format 4 Set up each parameters Clock Emissivity Change memory capability For t...

Страница 55: ...aser beam is not satisfied Replace the batteries The b symbol blinks or re install the batteries correctly The lens section is dirty Clean the filter referring to the Lens in Maintenance method of pag...

Страница 56: ...lling 2 Install new batteries in the correct direction The attached batteries are to be used for checking operations The battery life mentioned in Specifications is not assured for these batteries Pre...

Отзывы: