8 Important Information
8
8 Important Information
Please read the
instruction manual
completely before start-up and operation.
The latest version of the instruction manual applies.
Skin oil and contamination are absorbing in the UV spectral range. Finger
prints / contamination on the optically active surface are to be avoided. If
necessary, clean thoroughly with Isopropanol. Use gloves when cleaning.
The UVpad E is calibrated.
ATTENTION: Opening of the housing (except for the battery compartment)
leads to a loss of calibration.
Only use
batteries
, which are suitable for up to 60 °C.
The UVpad E is a highly sensitive optical measurement device. It must not be
exposed to shocks.
When using the UVpad E, light and UV radiation can be reflected and distributed
at the housing. Appropriate protection measures for the
protection against radi-
ation
need to be applied, if necessary.
The maximum housing temperature of the sensor can reach > 60 °C and there
is the risk of skin burns on contact. Please allow for the UVpad E to cool
down and use protective gloves to avoid
skin burns
.