9
Impulse Auto Mattress Instructions and Technical Specification
9.
In the case of emergency, pull hard to remove the mattress’ CPR latch and detach
the power unit by pressing the power/ mute button on the control panel at the same
time. The air will discharge from the mattress with patient’s own weight. To resume
normal operation, simply re-insert CPR latch securely, press the power/ mute button
again and reset the patient weight. See instructions beginning in Section 6.2.
10. Cleaning Instruction
The air mattress and power unit must be cleaned thoroughly between patients to
avoid cross contamination. The following is a suggested guideline. Be sure to follow
local infection control policies.
Regular cleaning can be performed at bedside with a disinfectant followed by
drying with a clean dry cloth. Use only mild detergents to clean the coverlet and the
mattress. Any appropriate NON-PHENOLIC cleaning agent may be used for heavy
and coverlet are completely dry before letting the patient lie on the surface again.
The recommended wash temperature is 70°C, at 95°C the shrinkage rate is 1%
Do not use electric or tumble dryers. Do not iron.
WARNING!
Always unplug the power unit before cleaning. Routine cleaning of
power unit can be done by wiping down with damp cloth using disinfectant or mild
detergent. Never spray liquids directly on the unit itself.
11. Routine Maintenance and
Storage
Only disinfected and dry systems should be stored. Disconnect the air hoses from
the power unit. Roll up the mattress starting from the head end and working down
toward the foot end. Use the straps to secure it and store in clean plastic or other
storage bag. Store in a cool, dark place.
Always replace the coverlet before using the mattress. If the coverlet is not properly
high risk of cross contamination if the mattress is used without a coverlet.
needed, contact
Care on 0333 222 8584.
Opera