background image

HOBO® Temp/RH 2.5% Data Logger (UX100-011) Manual 

16210-G 

The HOBO Temp/RH data logger records temperature and relative humidity (within 2.5% 
accuracy) in indoor environments with its integrated sensors. Using HOBOware®, you can easily 
configure the logger alarm to trip for specific high or low sensor readings. Or, you can set up burst 
logging in which the logger records data at a different interval during certain conditions. The 
logger can also calculate minimum, maximum, average, and standard deviation statistics. This 
compact data logger has a built-in LCD screen to monitor the current temperature, relative 
humidity, logging status, battery use, and memory consumption in between readouts. 

Specifications 

Temperature Sensor 

 

Range 

-20° to 70°C (-4° to 158°F) 

Accuracy 

±0.21°C from 0° to 50°C (±0.38°F from 32° to 122°F), see Plot A 

Resolution 

0.024°C at 25°C (0.04°F at 77°F), see Plot A 

Response Time 

4 minutes in air moving 1 m/s (2.2 mph) 

Drift 

<0.1°C (0.18°F) per year 

RH Sensor 

 

Range 

1% to 95% (non-condensing) 

Accuracy 

±2.5% from 10% to 90% typical to a maximum of ±3.5% including 
hysteresis, see Plot B 

Resolution 

0.05% at 25°C (77°F) 

Response Time 

11 seconds to 90% in airflow of 1 m/s (2.2 mph) 

Drift 

<1% per year typical 

Logger 

 

Logger Operating Range 

Logging: -20° to 70°C (-4° to 158°F); 0 to 95% RH (non-condensing) 
Launch/Readout: 0° to 50°C (32° to 122°F) per USB specification 

Logging Rate 

1 second to 18 hours, 12 minutes, 15 seconds 

Logging Modes 

Fixed interval (normal), burst, or statistics 

Memory Modes 

Wrap when full or stop when full 

Start Modes 

Immediate, push button, date & time, or next interval 

Stop Modes 

When memory full, push button, or date & time 

Restart Mode 

Push button 

Time Accuracy 

±1 minute per month at 25°C (77°F), see Plot C 

Power Source 

One 3V CR2032 lithium battery and USB cable 

Battery Life 

1 year, typical with logging rate of 1 minute and sampling interval of 
15 seconds or greater 

Memory 

128 KB (84,650 measurements, maximum) 

Download Type 

USB 2.0 interface 

Full Memory Download Time 

20 seconds 

LCD 

LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly 
or go blank in temperatures outside this range 

Size 

3.66 x 8.48 x 2.29 cm (1.44 x 3.34 x 0.9 in.) 

Weight 

30 g (1.06 oz) 

Environmental Rating 

IP50 

 

The CE Marking identifies this product as complying with all relevant 
directives in the European Union (EU). 

 

 

 

HOBO Temp/RH 2.5% 
Data Logger 

UX100-011 

Included Items: 

 

Command™ strip 

 

Double-sided tape 

 

Hook & loop strap 

Required Items: 

 

HOBOware 3.4 or later 

 

USB cable (included with 
software) 

 

Содержание UX100-011

Страница 1: ...to 90 in airflow of 1 m s 2 2 mph Drift 1 per year typical Logger Logger Operating Range Logging 20 to 70 C 4 to 158 F 0 to 95 RH non condensing Launch Readout 0 to 50 C 32 to 122 F per USB specificat...

Страница 2: ...ith an LCD screen that displays details about the current status This example shows all symbols illuminated on the LCD screen followed by definitions of each symbol in the table below LCD Symbol Descr...

Страница 3: ...setting up the logger as described in the next section When this option is enabled you can still temporarily view the LCD screen by pushing the Start Stop button for 1 second The LCD will then remain...

Страница 4: ...top button 9 Choose whether to keep the LCD on or off By default the LCD will always remain on while logging If you select the Turn LCD off checkbox the LCD will not show the current readings status o...

Страница 5: ...he logger would switch to burst logging mode again and record data every 10 seconds Once the temperature rises back to 32 F the logger will then return to normal mode logging every 5 minutes To set up...

Страница 6: ...g To set up statistics 1 Select Statistics for Logging Mode in the Launch Logger window If Statistics has already been configured for this logger click the Edit button in the Launch Logger window 2 Cl...

Страница 7: ...the L2 and L3 markings on the board shown here enlarged and without the logger case for illustrative purposes Hold the sensor at a slight angle as you are inserting it and let the pins come to rest in...

Страница 8: ...LCD should display HOBO briefly after correctly installing the battery WARNING Do not cut open incinerate heat above 85 C 185 F or recharge the lithium battery The battery may explode if the logger i...