to 75.999°F, burst logging can start when the sensor 
reading is 75.994°F (which is within the 0.02°C 
resolution). 

 

Once the high or low condition clears, the logging 
interval time will be calculated using the last recorded 
data point in burst logging mode, not the last data point 
recorded in “normal mode.” For example, let’s assume 
the logger has a 10-minute logging interval and logged a 
data point at 9:05. Then, the high limit was surpassed 
and burst logging began at 9:06. Burst logging then 
continued until 9:12 when the sensor reading fell back 
below the high limit. Now back in normal mode, the next 
logging interval will be 10 minutes from the last burst 
logging point, or 9:22 in this case. If burst logging had not 
occurred, the next data point would have been at 9:15. 

 

A New Interval event will appear on the plot (if you select 
events for plotting in the Plot Setup window) each time 
the logger enters or exits burst logging mode. 

Statistics 

Statistics is a logging mode in which the logger calculates 
maximum, minimum, average, and standard deviation statistics 
during logging, recording the results at each logging interval 
based on samples taken at a rate you specify. This will result in 
up to four additional series per sensor that record the following 
information at each logging interval: 

 

The maximum, or highest, sampled value, 

 

The minimum, or lowest, sampled value, 

 

An average of all sampled values, and 

 

The standard deviation from the average for all sampled 
values. 

For example, let’s say both the temperature and RH sensors 
have been enabled, the logging interval is set to 5 minutes and 
the sampling interval is set to 30 seconds (with maximum, 
minimum, average, and standard deviation all enabled). Once 
logging begins, the logger will measure and record the actual 
temperature and RH sensor values every 5 minutes. In addition, 
the logger will take a temperature and RH sample every 30 
seconds and temporarily store them in memory. The logger will 
then calculate the maximum, minimum, average, and standard 
deviation using the samples gathered over the previous 5-
minute period and log the resulting values. When reading out 
the logger, this would result in 10 data series (not including any 
derived series, such as dew point): two sensor series (with 
temperature and RH data logged every 5 minutes) plus eight 
maximum, minimum, average, and standard deviation series 
(four for temperature and four for RH with values calculated 
and logged every 5 minutes based on the 30-second sampling). 

To set up statistics: 

1.

 

Select Statistics for Logging Mode in the Launch Logger 
window. If Statistics has already been configured for this 
logger, click the Edit button in the Launch Logger window. 

2.

 

Click the Maximum, Minimum, Average, and Standard 
Deviation checkboxes for each of the statistics you want to 
calculate during logging. Note that Average is automatically 
enabled when selecting Standard Deviation. 

Important:

 

Statistics apply to all enabled sensors; every selected 
statistic will be calculated for all sensors (except battery 
voltage). For example, if both the temperature and RH 

sensors have been selected in the Launch Logger window 
and you select Average, then the average will be calculated 
for both temperature and RH. In addition, the more 
statistics you record, the shorter the logger duration and 
the more memory is required. 

 

3.

 

Set the sampling interval, which must be less than and a 
factor of the logging interval. Choose either a preset 
sampling interval or select Custom and enter your own 
sampling interval. Keep in mind that the more frequent the 
sampling rate, the greater the impact on battery life. 

4.

 

Click OK when done. This will return you to the Launch 
Logger window. Click the Edit button next to Logging Mode 
in the Launch Logger window to make additional changes. 

Once logging begins, click the Alarm/Stats button on the logger 
to cycle through the current maximum, minimum, average, and 
standard deviation data on the LCD screen. You can plot the 
statistics series once you read out the logger. 

Reading Out the Logger 

There are two options for reading out the logger: connect it to 
the computer with a USB cable and read out it with HOBOware, 
or connect it to a HOBO U-Shuttle (U-DT-1, firmware version 
1.18m030 or higher) and then offload the data files from the  
U-Shuttle to HOBOware. Refer to the HOBOware Help for more 
details. 

Recording Internal Logger Events 

The logger records the following internal events to track logger 
operation and status. You can plot these events in HOBOware 
after reading out the logger and opening the data file. 

Internal Event Name

Definition 

Host Connected 

The logger was connected to the computer. 

Started 

The Start/Stop button was pressed to begin 
logging. 

Stopped 

The logger received a command to stop 
recording data (from HOBOware or by 
pushing the Start/Stop button). 

Button Up/Button 
Down 

The Start/Stop button was pressed for 
1 second. 

Chan <#> Alarm 
Tripped 

A sensor alarm has tripped; # is the channel 
or sensor number as shown in the Launch 
Logger window in HOBOware. 

Chan <#> Alarm 
Cleared 

A sensor alarm has cleared; # is the channel 
or sensor number as shown in the Launch 
Logger window in HOBOware. This event also 
contains the value that was furthest out of 
range for the sensor before the alarm 
cleared. 

Содержание UX100-011

Страница 1: ...to 90 in airflow of 1 m s 2 2 mph Drift 1 per year typical Logger Logger Operating Range Logging 20 to 70 C 4 to 158 F 0 to 95 RH non condensing Launch Readout 0 to 50 C 32 to 122 F per USB specificat...

Страница 2: ...ith an LCD screen that displays details about the current status This example shows all symbols illuminated on the LCD screen followed by definitions of each symbol in the table below LCD Symbol Descr...

Страница 3: ...setting up the logger as described in the next section When this option is enabled you can still temporarily view the LCD screen by pushing the Start Stop button for 1 second The LCD will then remain...

Страница 4: ...top button 9 Choose whether to keep the LCD on or off By default the LCD will always remain on while logging If you select the Turn LCD off checkbox the LCD will not show the current readings status o...

Страница 5: ...he logger would switch to burst logging mode again and record data every 10 seconds Once the temperature rises back to 32 F the logger will then return to normal mode logging every 5 minutes To set up...

Страница 6: ...g To set up statistics 1 Select Statistics for Logging Mode in the Launch Logger window If Statistics has already been configured for this logger click the Edit button in the Launch Logger window 2 Cl...

Страница 7: ...the L2 and L3 markings on the board shown here enlarged and without the logger case for illustrative purposes Hold the sensor at a slight angle as you are inserting it and let the pins come to rest in...

Страница 8: ...LCD should display HOBO briefly after correctly installing the battery WARNING Do not cut open incinerate heat above 85 C 185 F or recharge the lithium battery The battery may explode if the logger i...

Отзывы: