InTemp CX405 RTD Dry Ice Logger Manual
1-508-759-9500 (U.S. and International) 4
www.onsetcomp.com
1-800-LOGGERS (U.S. only)
checks, connect to the logger and tap Perform (Morning,
Afternoon, or Daily) Check every day.
Once logging begins, the logger will display the current
temperature reading and the minimum and maximum readings
within the current 24-hour period (midnight from one day to
midnight the next day) if the logger was set to record logger
checks (see
Performing Logger Checks
). Otherwise the
minimum and maximum readings represent the entire logging
period and only reset when the logger is downloaded and
restarted or stopped and reconfigured (they will not reset if you
download the logger and continue logging). These minimum
and maximum values are also available in the logger reports
(see
Downloading the Loggers
). Click Next to switch between
external probe and ambient temperature readings. Minimum
and maximum readings are only available for the external
probe.
Logger Alarms
You can set up an alarm to trip on the logger when the
temperature reading falls above or below a specified value or
when the logger battery is low. Specific alarm settings are set in
the logger profile that you create in either InTempConnect or in
the app.
When a temperature alarm trips:
•
The logger LED will blink every 5 seconds.
•
The alarm icon will appear on the LCD and in the app.
•
The logger will beep once every 15 seconds (unless
audible alarms are disabled in the logger profile).
•
An Alarm Tripped event is logged.
When a low battery alarm trips:
•
The battery icon on the LCD will flash.
•
The logger will quickly beep three times every 15
seconds.
•
A Low Battery event is logged.
To mute a beeping alarm, press the Mute button on the logger.
Once muted, you cannot turn the beeping back on. Note that if
a temperature alarm is occurring at the same time as a low
battery alarm, pressing the mute button will silence both
alarms.
Download the logger to view details about a tripped alarm and
to clear temperature alarm indicators in the app and on the
LCD. Replace the batteries in the logger to clear a battery
alarm.
Note:
Download the logger before replacing the
batteries to ensure no data is lost.
Passkey Protection
The logger is protected by an encrypted passkey automatically
generated by the InTemp app for InTempConnect users and
optionally available if you are using the InTemp app only. The
passkey uses a proprietary encryption algorithm that changes
with every connection.
InTempConnect Users
Only InTempConnect users belonging to the same
InTempConnect account can connect to a logger once it is
configured. When an InTempConnect user first configures a
logger, it is locked with an encrypted passkey that is
automatically generated by the InTemp app. After the logger is
configured, only active users associated with that account will
be able to connect to it. If a user belongs to a different account,
that user will not be able to connect to the logger with the
InTemp app, which will display an invalid passkey message.
Administrators or users with the required privileges can also
view the passkey from the device configuration page in
InTempConnect and share them if needed. See
www.intempconnect.com/help for more details.
Note:
This
does not apply to InTempVerify. if the logger was configured
with a logger profile in which InTempVerify was enabled, then
anyone can download the logger with the InTempVerify app.
InTemp App Only Users
If you are using the InTemp app only (not logging in as an
InTempConnect user), you can create an encrypted passkey for
the logger that will be required if another phone or tablet
attempts to connect to it. This is recommended to ensure that
a deployed logger is not mistakenly stopped or purposely
altered by others.
To set a passkey:
1.
Tap the Devices icon and connect to the logger.
2.
Tap Set Logger Passkey.
3.
Type a passkey up to 10 characters.
4.
Tap Save.
5.
Tap Disconnect.
Only the phone or tablet used to set the passkey can then
connect to the logger without entering a passkey; all other
mobile devices will be required to enter the passkey. For
example, if you set the passkey for the logger with your tablet
and then try to connect to the device later with your phone,
you will be required to enter the passkey on the phone but not
with your tablet. Similarly, if others attempt to connect to the
logger with different devices, then they would also be required
to enter the passkey. To reset a passkey, simultaneously press
both the top and bottom button on the logger for 5 seconds, or
connect to the logger, tap Set Logger Passkey, and select Reset
Passkey to Factory Default.
Downloading the Logger
You can download the logger to a phone or tablet and generate
reports that include logged probe and ambient readings,
events, user activity, alarm information, and more. Reports can
be shared immediately upon download or accessed later in the
InTemp app.
InTempConnect users:
Privileges are required to download,
preview, and share reports in the InTemp app. Report data is
automatically uploaded to InTempConnect when you download
the logger. Log into InTempConnect to build custom reports
(requires privileges). In addition, InTempConnect users can also
download CX loggers automatically on a regular basis using the
CX5000 Gateway. Or, if the logger was configured with a logger
profile in which InTempVerify was enabled, then anyone can
download the logger with the InTempVerify app. For details on
the gateway and InTempVerify, see www.intempconnect/help.
To download the logger with the InTemp app:
1.
Tap the Devices icon and connect to the logger.
2.
Tap Download.
3.
Choose a download option: