background image

 

HOBO MX Temp/RH Data Logger (MX1101) Manual 

 

1-800-LOGGERS 6 

www.onsetcomp.com 

audible alarm is cleared, a visual alarm may remain on 
the logger LCD and in the app depending on the settings 
selected for maintaining visual alarms or because the 
alarm condition may still be in effect. In addition, an 
audible alarm will continue beeping when the sensor 
values have returned to the normal range until it is 
cleared as described in step 8. 

 

Although an audible alarm and a visual alarm can occur 
at the same time when a sensor alarm is tripped, they are 
cleared in different ways. The audible alarm can be 
cleared as described in step 8. Meanwhile, a visual alarm 
is cleared as determined by the setting selected for 
Maintain Visual Alarm Until configuration setting. This 
means you could clear a beeping audible alarm and the 
visual alarm will remain on the LCD and in the app until 
the logger is reconfigured, the sensor is in limits, or the 
alarm button is pressed--whichever setting you selected. 

 

If the logger was configured to stop logging with a button 
push, any tripped alarms will be cleared automatically 
when logging is stopped and no Alarm Cleared event will 
be logged. This ensures that the logger will start checking 
for alarm conditions when logging resumes (if the logger 
was configured with Allow Button Restart selected). 

Burst Logging 

Burst logging is a logging mode that allows you to set up more 
frequent logging when a specified condition is met. For 
example, a logger is recording data at a 5-minute logging 
interval and burst logging is configured to log every 30 seconds 
when the temperature rises above 85°F (the high limit) or falls 
below 32°F (the low limit). This means the logger will record 
data every 5 minutes as long as the temperature remains 
between 85°F and 32°F. Once the temperature rises above 
85°F, the logger will switch to the faster logging rate and record 
data every 30 seconds until the temperature falls back to 85°F. 
At that time, logging then resumes every 5 minutes at the 
normal logging interval. Similarly, if the temperature falls below 
32°F, then the logger would switch to burst logging mode again 
and record data every 30 seconds. Once the temperature rises 
back to 32°F, the logger will then return to normal mode, 
logging every 5 minutes. 

Note:

 Sensor alarms, statistics, and the 

Stop Logging option “Wrap When Full” are not available in 
burst logging mode. 

To set up burst logging: 

1.

 

Tap Devices. Tap the logger in the app to connect to it and 

tap . 

2.

 

Tap Logging Mode and then tap Burst Logging. 

3.

 

Select Low and/or High and either type or drag the slider to 
set the low and/or high values. 

4.

 

Repeat step 3 for the other sensor if desired. 

5.

 

Set the burst logging interval, which must be faster than the 
logging interval. Keep in mind that the more frequent the 
burst logging rate, the greater the impact on battery life 
and the shorter the logging duration. 

6.

 

Tap Save. 

7.

 

Tap . 

Notes: 

 

The high and low burst limits are only checked when the 
logger's LCD screen refreshes once every 15 seconds. 
Therefore, if you set the logging interval to less than 15 
seconds and the sensor reading falls outside the levels, 
the burst logging will not begin until the next 15-second 
refresh cycle. 

 

If high and/or low limits have been configured for more 
than one sensor, then burst logging will begin when any 
high or low condition goes out of range. Burst logging will 
not end until all conditions on all sensors are back within 
normal range. 

 

The actual values for the burst logging limits are set to 
the closest value supported by the logger. 

 

Burst logging mode can begin or end when the sensor 
reading is within the resolution specifications. This means 
the value that triggers burst logging may differ slightly 
than the value entered.  

 

Once the high or low condition clears, the logging 
interval time will be calculated using the last recorded 
data point in burst logging mode, not the last data point 
recorded in normal mode. For example, a logger has a 
10-minute logging interval and logged a data point at 
9:05. Then, the high limit was surpassed and burst 
logging began at 9:06. Burst logging then continued until 
9:12 when the sensor reading fell back below the high 
limit. Now back in normal mode, the next logging interval 
will be 10 minutes from the last burst logging point, or 
9:22 in this case. If burst logging had not occurred, the 
next data point would have been at 9:15. 

 

A New Interval event is created each time the logger 
enters or exits burst logging mode. See 

Logger Events

 for 

details on plotting and viewing the event. In addition, if 
the logger is stopped with a button push while in burst 
logging mode, then a New Interval event is automatically 
logged and the burst condition is cleared, even if the 
actual high or low condition has not cleared. The logger 
will check the high and low conditions when logging 
resumes (if the logger was configured with Allow Button 
Restart selected). 

Statistics Logging 

During fixed logging, the logger records data for enabled 
sensors and/or selected statistics at the logging interval 
selected. Statistics are calculated at a sampling rate you specify 
with the results for the sampling period recorded at each 
logging interval. The following statistics can be logged for each 
sensor: 

 

The maximum, or highest, sampled value, 

 

The minimum, or lowest, sampled value, 

 

An average of all sampled values, and 

 

The standard deviation from the average for all sampled 
values. 

For example, a logger is configured with both the temperature 
and RH sensors enabled, and the logging interval set to 5 
minutes. The logging mode is set to fixed logging with Normal 
and all four statistics enabled and with a statistics sampling 
interval of 30 seconds. Once logging begins, the logger will 
measure and record the actual temperature and RH sensor 

Содержание HOBO MX1101

Страница 1: ...Temperature 7 30 minutes in air moving 1 m s 2 2 mph RH 20 seconds to 90 in airflow of 1 m s 2 2 mph Logger Radio Power 1 mW 0 dBm Transmission Range Approximately 30 5 m 100 ft line of sight Wireles...

Страница 2: ...to reset a logger password Alarm Stats Button Press this button for 1 second to switch between statistics alarm readings and the current sensor readings as applicable or to silence a beeping alarm Pre...

Страница 3: ...in to cycle through any statistics defined above and ultimately back to the current sensor reading This is an example of a temperature reading Temperature units are determined by the settings in the a...

Страница 4: ...used as the name 4 Tap Group to add the logger to a group optional Tap Save 5 Tap Logging Interval and choose how frequently the logger will record data unless operating in burst logging mode see Burs...

Страница 5: ...or reading rises above the high alarm value Drag the slider or type a value to set the high alarm value 4 Select Low if you want an alarm to trip when the sensor reading falls below the low alarm valu...

Страница 6: ...he burst logging rate the greater the impact on battery life and the shorter the logging duration 6 Tap Save 7 Tap Notes The high and low burst limits are only checked when the logger s LCD screen ref...

Страница 7: ...nother phone or tablet attempts to connect to it This is recommended to ensure that a deployed logger is not mistakenly stopped or purposely altered by others This password uses a proprietary encrypti...

Страница 8: ...the logger operating range Expected battery life varies based on the ambient temperature where the logger is deployed the logging or sampling interval frequency of offloading and connection to the mo...

Страница 9: ...and 2 this device must accept any interference received including interference that may cause undesired operation FCC Caution Any changes or modifications not expressly approved by the party responsi...

Отзывы: