background image

 

15 

 

Sample is not Stable 

Improve the sample 

The concentration of sample is 

too high 

Diluted sample 

Power  Supply  Voltage  Low  or 

not Stable 

Improve the Power Supply 

Lamp  damage  or  lamp  life 

maturity 

Replace lamp 

Dark  Current  Error  when  self-

check 

The  lid  of  the  compartment  is 

open during self-check

 

Close the lid, restart 

System Calibrate Failed 

Something block the Light path  Remove it, calibrate again 

Power on, back light is OK, but 

nothing display on the screen or 

display is not clear 

Display Contrast problem 

Adjust 

the 

contrast 

potentiometer 

Measurements inaccurate 

Cuvettes were contaminated 

Clean cuvettes 

Samples were contaminated 

Improve samples 

Worse matching of the cuvettes 

Improve  the  matching  of  the 

cuvettes 

Dark current error 

Resample dark current 

Repair and Maintenance 

1.  Daily Maintain 

Check the Compartment 

After measurement, the cuvettes with sample solutions should be taken out of the compartment 

in time. Or the volatilization of the solution would make the mirror go moldy. Users must pay 

more attention to the corrosive sample and liquid easy to volatilize. Any solution remains in the 

compartment should be wiped off immediately. 

Surface Clean 

The  cover  of  the  instrument  is  with  paint.  Please  use  wet  towel  to  wipe  off  the  drips  on  the 

surface immediately. Organic solution is forbidden to be used to clean the cover. Please wipe 

off the dirt on the cover timely. 

Clean the Cuvettes 

After  every  test  or  after  a  solution  change,  the  cuvettes  should  be  cleaned  carefully,  or  the 

remains on the surface would cause measuring error. 

2.  Spare Parts Replacement 

Replace the Fuse 

Содержание V-10 PLUS

Страница 1: ...VISIBLE SPECTROPHOTOMETER V 10 PLUS User s Manual...

Страница 2: ...ls 2 Main Specifications 3 Description of Appearance and Keys 4 1 Appearance 4 2 Keypad 5 3 Description of Keys 5 Functions 6 Getting Started 6 Important Guidelines 7 General Operating 7 Measuring 8 1...

Страница 3: ...II Warranty 19 Equipment Disposal 19...

Страница 4: ...om us and authorized contractual partners The device must be connected to a power outlet that has a protective ground connection If the equipment is used in a manner not specified by the manufacturer...

Страница 5: ...ded outlet If the local voltage is not stable a voltage regulator is required Be away from direct sunlight 2 Install Spectrophotometer Placement Place the instrument on the stable table carefully Inst...

Страница 6: ...325 1000nm Wavelength Accuracy 2nm Wavelength Repeatability 0 8nm Photometric Range 0 3 3A 0 200 T Photometric Accuracy 0 5 T Photometric Repeatability 0 3 T Spectral Bandwidth 4nm Stray Light 0 3 T 3...

Страница 7: ...Keys 1 Appearance Front View 1 2 3 4 Back View 5 10 9 8 7 6 11 1 LCD Display 2 Keypad 3 Lid of Sample Room 4 Rod 5 LCD Contrast Adjust 6 Printer port 7 USB port 8 Cover of Fan 9 Power Socket 10 Power...

Страница 8: ...5 2 Keypad 3 Description of Keys SET SET Key Set Parameters GOTO GOTO Key Set Wavelength ZERO ZERO Key Blank PRINT PRINT Key Print measuring result...

Страница 9: ...owing chart describes the basic operation of the instrument Turn On and Self check Switch on the power Then the instrument begins to self check and 20 minutes warm up Self check includes the following...

Страница 10: ...uring solutions and cleaning and disinfectant materials in accordance with the relevant local laboratory regulations General Operating Select Application Main interface press the key SELECT left to en...

Страница 11: ...the Standard Curve After got the Standard Curve by Marked press input the file name and press SELECT left to save Load the Standard Curve Quantitative interface press to select Load Curve press to ch...

Страница 12: ...LECT left to record W L 500 0nm 0 000A No W L T Abs 1 500 0 100 0 0 000 2 500 0 100 0 0 000 Test Cancel 2 Quantitation Step 1 Start Quantitation Main Interface press key SELECT left to choose Quantita...

Страница 13: ...cient K and B Press to input coefficient K press SELECT left same way set B W L 500 0nm W or k W L 500 0 nm O K Cancel W L 500 0nm Coef f i ci ent K 00000 O K Cancel Method 2 Establish Standard Curve...

Страница 14: ...ndard samples in the light path as the screen indicates press to input the concentration press SELECT left to confirm to finish all the standard samples W L 500 0nm Num ber 3 O K Cancel W L 500 0nm I...

Страница 15: ...es Put the sample to be tested in the light path press SELECT left to measure Then the test result will display in the data sheet Repeat this step to finish measuring all the samples 500 0nm 0 000A No...

Страница 16: ...Test M ode D2 Lam p O n W Lam p O n O K Cancel O K Cancel Turn On Off W Lamp Press to choose W Lamp then press SELECT left to enter into Press to choose On or Off press SELECT left to turn on off W L...

Страница 17: ...ut Version Press to choose Version press SELECT left to view version information press any key to return Sof t war e V2 4 2 Har dwar e V1 2 Dar k Cur r ent Reset W L Ver si on O K Cancel Troubleshooti...

Страница 18: ...he matching of the cuvettes Dark current error Resample dark current Repair and Maintenance 1 Daily Maintain Check the Compartment After measurement the cuvettes with sample solutions should be taken...

Страница 19: ...er supply Switch off the power supply and unplug the socket Step 3 Take out the Fuse Seat Take out the fuse seat by the screwdriver Step 4 Replace a new fuse Pick out the spare fuse 3 15A 250V and rep...

Страница 20: ...he socket Step 3 Open the cover Unscrew the 4 screws indicated Each side with 2 screws and remove the cover Rod Step 4 Open the cover of the light chamber Unscrew the 2 screws on the light chamber cov...

Страница 21: ...move the lamp seat to Left or Right until it focus on the center of the slot Then fix the screws If the facular deviate to Up and Down then loosen the No 2 screws and move the lamp seat Up and Down u...

Страница 22: ...ARRANTY OF MERCHANTABILITY Equipment Disposal This equipment is marked with the crossed out wheeled bin symbol to indicate that this equipment must not be disposed of with unsorted waste Instead it s...

Страница 23: ......

Страница 24: ...Giorgio Bormac S r l Add VIA DELLA MECCANICA 25 41012 CARPI MO ITALY Tel 39 059 653 274 Fax 39 059 653 282...