On-X

®

 Prosthetic Heart Valve | Instructions for Use 

8

Instrument Handle

The instrument handle (Figure 4) facilitates holding the 
valve or the rotator during surgery. The instrument handle 
is comprised of a grip, a bendable stem, and a tip.

Rotator

The valve rotator (Figure 5) is used for reorienting an in 
situ valve and may be used to verify leaflet mobility. The 
rotator consists of a plastic head with a centrally located 
leaflet probe and an attached handle. 
The rotator may be used with or without the instrument 
handle attached. To attach the rotator to the instrument 
handle, insert the instrument handle tip directly into the 
slot on the end of the rotator handle. The rotator snaps 
into place after the application of a light insertion force.

Leaflet Probe 

The leaflet probe (Figure 6) is a flexible rod with tapered 
ends. The leaflet probe may be used to gently move the 
leaflets to verify that they open and close freely.

WARNING: DO NOT sterilize instruments with any 
method of sterilization other than steam. Damage 
to some items could result from use of other 
sterilization methods.
WARNING: The rotator must be removed from 
the handle after use and prior to cleaning. A force 
greater than the insertion force is required to 
remove the rotator from the instrument handle.

9.  DIRECTIONS FOR USE
WARNING: DO NOT use the On-X Prosthetic Heart 
Valve if:
•  the prosthesis has been dropped, damaged, or 

mishandled in any way;

•  the expiration date has elapsed;
•  the tamper evident seal is broken;
•  the serial number tag does not match the serial 

number on the container label.

9.1  Physician Training

No special training is required to implant the On-X 
Prosthetic Heart Valve. The techniques for implanting 
this prosthesis are similar to those used for any 
mechanical heart valve prosthesis.

9.2  Sterilization and Resterilization

The On-X Prosthetic Heart Valve is provided sterile. 
If the sterility expiration date has passed or if upon 
removal from the outer box, the valve container is 
damaged or the sterility barrier is broken, do not use 
the valve. Call On-XLTI Customer Service and arrange to 
return the valve and receive a replacement.

WARNING: If during surgery the valve is removed 
from its container but not used, it must not be 
repackaged or resterilized. 

In this situation, the valve 

must be returned to On-XLTI. Call Customer Service for 
information before any return is made.

WARNING: Do not resterilize the On-X Prosthetic 
Heart Valve.

9.3  Handling and Preparation Instructions 
CAUTION: Handle the prosthesis with only On-XLTI 
On-X Prosthetic Heart Valve Instruments. Only On-
XLTI On-X Prosthetic Heart Valve sizers should be 
used during the selection of the valve size; other 
sizers may result in improper valve selection.
CAUTION: Avoid contacting the carbon surfaces 
of the valve with gloved fingers or any metallic or 
abrasive instruments as they may cause damage to 
the valve surface not seen with the unaided eye that 
may lead to accelerated valve structural dysfunction, 
leaflet escape, or serve as a nidus for thrombus 
formation.
CAUTION: Avoid damaging the prosthesis through 
the application of excessive force to the valve orifice 
or leaflets.

Figure 4: Instrument Handle 

Figure 5: Rotator

Figure 6: Leaflet Probe 

8.5  Accessory Cleaning and Sterilization

Instruments for implantation of the On-X Prosthetic 
Heart Valve are supplied separately, NON-STERILE, and 
must be cleaned and sterilized prior to use. Standard 
hospital surgical instrument cleaning procedures 
must be used. Note: the metallic instruments are 
made of titanium. The plastic instruments are made of 
polyphenylsulfone. Materials used in these instruments 
can withstand standard steam and flash steam 
sterilization.

WARNING: These instruments are NOT provided 
sterile. They must be properly cleaned and sterilized 
prior to each use.

Содержание On-X Aortic Heart Valve and Extended Holder

Страница 1: ...eart Valve with Conform X Sewing Ring REF ONXAC On X Mitral Heart Valve with Conform X Sewing Ring REF ONXMC On X Aortic Heart Valve with Anatomic Sewing Ring REF ONXAN On X Aortic Heart Valve and Ext...

Страница 2: ...SSMENT AND VALVE ROTATION 11 9 7 VALVE ORIENTATION 12 10 POSTOPERATIVE INFORMATION 12 10 1 MAGNETIC RESONANCE IMAGING MRI COMPATIBILITY 12 10 2 RETURNED GOODS 12 11 PATIENT INFORMATION 12 11 1 PATIENT...

Страница 3: ...eplica sizers verify the aortic valve 10 Figure 12 Supra annular valve positioning 11 Figure 13 Sewing ring cross sections 11 Figure 14 Removing the valve holder 11 Figure 15 Insert valve rotator 12 F...

Страница 4: ...a travel arc of 50 to the closed position The orifice is composed of graphite substrate coated with On X Carbon a pure unalloyed form of pyrolytic carbon The leaflets consist of On X Carbon deposited...

Страница 5: ...esis regurgitation prosthesis structural dysfunction prosthesis thrombosis stroke thromboembolism It is possible that these complications could lead to reoperation explantation permanent disability de...

Страница 6: ...rded on the outer package label Appropriate inventory control should be maintained so that prostheses with earlier expiration dates are preferentially implanted and expiration is avoided To protect th...

Страница 7: ...YES Intrasupra annular 23 Aortic Cylindrical YES Intrasupra annular 25 Aortic Cylindrical YES Intrasupra annular 27 29 Aortic Conical YES Intra annular 19 Aortic Conform X Cylindrical YES Intrasupra...

Страница 8: ...n and Resterilization The On X Prosthetic Heart Valve is provided sterile If the sterility expiration date has passed or if upon removal from the outer box the valve container is damaged or the steril...

Страница 9: ...the patient or place it in the patient s records Figure 7a Twist off outer lid Figure 7b Remove by pull tab Figure 7c or invert on sterile field 4 Open the outer container Twist off outer lid package...

Страница 10: ...thetic Heart Valve sizers when sizing the annulus Sizers contain cylindrical conical and aortic replica ends Refer to Table 2 to facilitate sizer selection Cylindrical sizers correspond to the valve s...

Страница 11: ...mplanted using a pledgetted or non pledgetted everting mattress suture technique although non everting and continuous suture techniques have also been used with success CAUTION When seating the valve...

Страница 12: ...e to rotation is encountered The torque required to rotate the valve in situ should be about the same as that required when testing rotation before implantation If noticeably greater torque is require...

Страница 13: ...mplete the card and carry it with them at all times 11 3 Patient Information Booklet On XLTI has made available a patient information booklet that the physician may choose to provide to the patient pr...

Страница 14: ...s in an international multicenter prospective non randomized study with retrospective controls Patients requiring isolated mitral heart valve replacement were enrolled from 1996 to 2001 at 16 centers...

Страница 15: ...d as linearized rates based on total patient years 3 Freedom from event was calculated based on the method of Kaplan Meier SE Standard Error 4 n number of patients in each category N total number of s...

Страница 16: ...3 0 7 99 0 0 7 99 0 0 7 Explant 1 0 4 3 0 7 98 0 1 0 98 0 1 0 Hemolysis5 0 0 0 0 0 0 100 0 0 100 0 0 Hemorrhage6 all 4 1 8 6 1 4 96 4 1 3 94 4 2 0 Hemorrhage major 4 1 8 2 0 5 97 0 1 2 97 0 1 2 Periv...

Страница 17: ...tis 18 7 8 Gastroinestinal Ulcer 18 7 8 Chronic Kidney Failure 13 5 6 Carotid Artery Disease 12 5 2 Coronary Artery Bypass Graft 10 4 4 Cancer 10 4 4 Previous Mitral Valve Replacement 9 3 9 Cardiomyop...

Страница 18: ...ts followed reproduced from Figure 2 Nd number of patients for which NYHA data were collected not including missing 3 n number of patients in each category 4 Undetermined means data were collected but...

Страница 19: ...2 0 4 3 Regurgitation Nd 20 Nd 37 Nd 68 Nd 36 Nd 25 n n Nd n n Nd n n Nd N n Nd n n Nd 0 5 25 0 9 24 3 27 39 7 17 47 2 7 28 0 1 2 12 60 0 25 67 6 37 54 4 16 44 4 17 68 0 3 2 10 0 0 0 0 3 4 4 2 5 6 1 4...

Страница 20: ...Bldg B Austin Texas 78752 U S A TEL 512 339 8000 FAX 512 339 3636 WEB www onxlti com EMAIL onx onxlti com CryoLife Europa Ltd Bramley House The Guildway Old Portsmouth Road Guildford Surrey GU3 1LR Un...

Отзывы: