On-X On-X Aortic Heart Valve and Extended Holder Скачать руководство пользователя страница 10

English

On-X

®

 Prosthetic Heart Valve | Instructions for Use 

10

Scrub Nurse/Surgeon:

1.  Check the tamper evident seal of the inner container.

  WARNING: DO NOT use the On-X Prosthetic Heart 

Valve if the tamper evident seal has been broken. 
If the tamper evident seal has been broken, use 
another valve and arrange for return through On-
XLTI Customer Service.

2.  Open the inner container by gently twisting the lid 

to break the tamper-proof seals (Figure 9) and then 
lifting the lid off the base. 

3.  Press the instrument handle tip into the slot on the 

valve holder until it snaps firmly into position (Figure 
10). Gently lift the valve out of the container and slide 
the holder plate off the holder. 

  Carefully grasp the sewing ring with a gloved hand 

using a light grip and gently turn the instrument 
handle in either direction. The valve should easily 
rotate within the sewing ring. Stop rotation testing 
with an orientation mark aligned with the pivot axis. 

  WARNING: DO NOT use the On-X Prosthetic 

Heart Valve if the valve does not rotate easily. Use 
another valve and arrange for return through On-
XLTI Customer Service.

6.  The valve is now ready for implantation. To ease 

positioning during implantation, the instrument 
handle stem can be bent by grasping the ends of the 
handle and the stem, then bending. Avoid grasping 
the valve.

  WARNING: DO NOT use the valve for leverage in 

bending the instrument handle. This could damage 
the valve and lead to mechanical failure.

9.4  Device Implantation 
WARNING: All accessory instruments must be 
cleaned and sterilized prior to use according to the 
instrument instructions. 

Sizing

Use only On-X Prosthetic Heart Valve sizers when sizing 
the annulus. Sizers contain cylindrical, conical, and aortic 
replica ends. Refer to Table 2 to facilitate sizer selection.
Cylindrical sizers correspond to the valve sizes 19 mm 
through 25 mm. Conical sizers correspond to the valve 
size 27/29 mm and 31/33 mm. These types of sizers may 
be used for both aortic and mitral valves.
The correct valve size is determined by obtaining a 
comfortable, not tight, fit of the sizer within the annulus. 
When a comfortable fit is found, the corresponding 
valve size is signified by the identification on the sizer. 
On-X Mitral Conform-X Prosthetic Heart Valves may be 
used when a comfortable fit is at or between size 25 and 
size 33. 
Aortic replica sizers are provided for all aortic valve sizes. 
For size 19 mm through 25 mm aortic valves, the aortic 
replica sizers are used to verify that the aortic valve can 
be properly seated in the annulus and that the coronary 
arteries remain unobstructed. Size 19 mm through 25 
mm aortic valves of standard, Conform-X, and Anatomic 
sewing ring configurations are designed to fit within the 
annulus at implant such that the exposed carbon flare 
rests in the annulus and the sewing ring is intrasupra-
annular (Figure 11).

Figure 11. Aortic replica sizers verify the aortic valve

Figure 10. Inserting the 
instrument handle

Figure 9. Opening the inner container

4.  Check the serial number tag against the label on the 

outer container.

  WARNING: DO NOT use the On-X Prosthetic Heart 

Valve if the serial number tag does not match the 
container label. Use another valve and arrange for 
a return through On-XLTI Customer Service.

5.  Remove the serial number tag by cutting the suture 

that holds it on the valve. If desired, the tag can 
be used to check for sterility by standard culture 
techniques immediately after it is removed.

Содержание On-X Aortic Heart Valve and Extended Holder

Страница 1: ...eart Valve with Conform X Sewing Ring REF ONXAC On X Mitral Heart Valve with Conform X Sewing Ring REF ONXMC On X Aortic Heart Valve with Anatomic Sewing Ring REF ONXAN On X Aortic Heart Valve and Ext...

Страница 2: ...SSMENT AND VALVE ROTATION 11 9 7 VALVE ORIENTATION 12 10 POSTOPERATIVE INFORMATION 12 10 1 MAGNETIC RESONANCE IMAGING MRI COMPATIBILITY 12 10 2 RETURNED GOODS 12 11 PATIENT INFORMATION 12 11 1 PATIENT...

Страница 3: ...eplica sizers verify the aortic valve 10 Figure 12 Supra annular valve positioning 11 Figure 13 Sewing ring cross sections 11 Figure 14 Removing the valve holder 11 Figure 15 Insert valve rotator 12 F...

Страница 4: ...a travel arc of 50 to the closed position The orifice is composed of graphite substrate coated with On X Carbon a pure unalloyed form of pyrolytic carbon The leaflets consist of On X Carbon deposited...

Страница 5: ...esis regurgitation prosthesis structural dysfunction prosthesis thrombosis stroke thromboembolism It is possible that these complications could lead to reoperation explantation permanent disability de...

Страница 6: ...rded on the outer package label Appropriate inventory control should be maintained so that prostheses with earlier expiration dates are preferentially implanted and expiration is avoided To protect th...

Страница 7: ...YES Intrasupra annular 23 Aortic Cylindrical YES Intrasupra annular 25 Aortic Cylindrical YES Intrasupra annular 27 29 Aortic Conical YES Intra annular 19 Aortic Conform X Cylindrical YES Intrasupra...

Страница 8: ...n and Resterilization The On X Prosthetic Heart Valve is provided sterile If the sterility expiration date has passed or if upon removal from the outer box the valve container is damaged or the steril...

Страница 9: ...the patient or place it in the patient s records Figure 7a Twist off outer lid Figure 7b Remove by pull tab Figure 7c or invert on sterile field 4 Open the outer container Twist off outer lid package...

Страница 10: ...thetic Heart Valve sizers when sizing the annulus Sizers contain cylindrical conical and aortic replica ends Refer to Table 2 to facilitate sizer selection Cylindrical sizers correspond to the valve s...

Страница 11: ...mplanted using a pledgetted or non pledgetted everting mattress suture technique although non everting and continuous suture techniques have also been used with success CAUTION When seating the valve...

Страница 12: ...e to rotation is encountered The torque required to rotate the valve in situ should be about the same as that required when testing rotation before implantation If noticeably greater torque is require...

Страница 13: ...mplete the card and carry it with them at all times 11 3 Patient Information Booklet On XLTI has made available a patient information booklet that the physician may choose to provide to the patient pr...

Страница 14: ...s in an international multicenter prospective non randomized study with retrospective controls Patients requiring isolated mitral heart valve replacement were enrolled from 1996 to 2001 at 16 centers...

Страница 15: ...d as linearized rates based on total patient years 3 Freedom from event was calculated based on the method of Kaplan Meier SE Standard Error 4 n number of patients in each category N total number of s...

Страница 16: ...3 0 7 99 0 0 7 99 0 0 7 Explant 1 0 4 3 0 7 98 0 1 0 98 0 1 0 Hemolysis5 0 0 0 0 0 0 100 0 0 100 0 0 Hemorrhage6 all 4 1 8 6 1 4 96 4 1 3 94 4 2 0 Hemorrhage major 4 1 8 2 0 5 97 0 1 2 97 0 1 2 Periv...

Страница 17: ...tis 18 7 8 Gastroinestinal Ulcer 18 7 8 Chronic Kidney Failure 13 5 6 Carotid Artery Disease 12 5 2 Coronary Artery Bypass Graft 10 4 4 Cancer 10 4 4 Previous Mitral Valve Replacement 9 3 9 Cardiomyop...

Страница 18: ...ts followed reproduced from Figure 2 Nd number of patients for which NYHA data were collected not including missing 3 n number of patients in each category 4 Undetermined means data were collected but...

Страница 19: ...2 0 4 3 Regurgitation Nd 20 Nd 37 Nd 68 Nd 36 Nd 25 n n Nd n n Nd n n Nd N n Nd n n Nd 0 5 25 0 9 24 3 27 39 7 17 47 2 7 28 0 1 2 12 60 0 25 67 6 37 54 4 16 44 4 17 68 0 3 2 10 0 0 0 0 3 4 4 2 5 6 1 4...

Страница 20: ...Bldg B Austin Texas 78752 U S A TEL 512 339 8000 FAX 512 339 3636 WEB www onxlti com EMAIL onx onxlti com CryoLife Europa Ltd Bramley House The Guildway Old Portsmouth Road Guildford Surrey GU3 1LR Un...

Отзывы: