OMS ARCADIA EXT S.T.01/3R ED.0 REV.8 09/2020
28
3.4.3.
ARCADIA EXT COMBINE TO DENTAL UNITS WITH SOSPENDED WATER GROUP
VIRTUOSUS CLASSIC
The operation of the controls is as follows:
i.
Rinse command (push the button towards the outside): starting from any position, this
command activates the ascent of backrest allow the patient to enter the glass rinsing; If the
command is pressed again the backrest returns to its previous position in the rinse cycle.
ii.
Reset command/ memory recall 1 (push the button towards the inside): ): starting from any
position other than the one reset, this command activates the descent of the patient chair
pantograph in the minimum position and simultaneously the backrest up to the maximum of
its travel; When the chair is to the zero position, pressing the button activates the memory
recall 1 of the chair.
E.
Selected program in instruments table
F.
Not available
PROGRAM STORAGE
To store the channel position simply manually move the chair to the desired position (using the manual
controls) and then press and hold the button for the program that you want to memorize (F, G or H) until you
hear the beep confirming
Содержание ARCADIA EXT
Страница 1: ...Instructions for use Patient Chair ARCADIA EXT...
Страница 11: ...OMS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 11 Do not sit at the end of the head or feet of the chair...
Страница 25: ...OMS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 25...
Страница 31: ...OMS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 31...
Страница 33: ...OMS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 33...
Страница 35: ...OMS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 35...
Страница 40: ...OMS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 40...
Страница 44: ...OMS ARCADIA EXT S T 01 3R ED 0 REV 8 09 2020 44...
Страница 63: ......