Setting Output Processing of Sensing Information
■
Setting Two-point Scaling
Measurement is performed at two positions and offset values are set for those mea-
sured values.
Example:
Correcting display values to match actual distances
Separate the two specified points by at least 1 % of the rated measurement range for the connected
Sensor Head.
Example: For the ZS-LD80 (diffuse reflection type)
The two measured points must be separated by at least “30 mm × 0.01 = 0.3 mm” as the measuring
range is “30 mm ±15 mm”.
4.0
-3.0
-2.8
3.8
-3.0
4.0
Display value
(mm)
Measured
value (mm)
Sensor Head
Actual distance:4.0 mm
Display value:3.8 mm
Sensing
object
Sensing
object
Sensor Head
Actual distance:-3.0 mm
Display value:-2.8 mm