background image

To View the Readings Stored in Memory

1.

Press the 

 button, while the average value is 

displayed.

The Memory number 
appears for a second 
before the pulse rate is 
displayed. The newest 
set is numbered “1”.

2.

Press the 

 button repeatedly to view the 

readings stored in memory.

To View the Weekly Averages

This unit calculates and displays weekly averages for the 
measurements taken in the morning and evening within 8 
weeks for each user.

Note: The week begins Sunday at 2:00.

1.

Select your User ID.

2.

Press the 

 button once.

The morning average for the current 
week “THIS WEEK” appears on the 
display.

3.

Press the 

 button again.

The evening average for the current 
week “THIS WEEK” appears on the 
display.

4.

Continue to press the 

 button to display the 

previous weeks.

The unit displays “-1 WEEK” for the previous week to 
“-7 WEEK” for the oldest set of averages. 

5.

Press the START/STOP button to turn off the 
monitor.

Morning Weekly Average

This is the average for the measurements taken during the 
morning (4:00 - 11:59) between Sunday and Saturday. An 
average for each day is calculated for up to three 
measurements taken within 10 minutes of the first 
measurement of the morning.

Evening Weekly Average 

This is the average for the measurements taken during the 
evening (19:00 - 1:59) between Sunday and Saturday. An 
average for each day is calculated for up to three 
measurements taken within 10 minutes of the last 
measurement of the evening.

The Blood Pressure Trend Graph

This unit also displays the Blood Pressure Trend Graph which 
shows the weekly averages for the Systolic Blood Pressure 
(SYS) taken in the morning and evening within 8 weeks for 
each user. The graph is displayed 4 weeks averages so that 
you can see the continuous blood pressure change.

To Delete All the Values Stored in Memory

The values stored in the memory are deleted by 
User ID.

When the memory symbol (

) appears, first press 

the 

 button. Then while holding it down, press the 

START/STOP button simultaneously for about 2 - 3 
seconds.

Note: You cannot partially delete the stored readings.

All values for the user you select will be deleted.

4. Troubleshooting and Maintenance

4.1 The Icons and Error Messages

Note: The irregular heartbeat symbol (

) and the movement error 

symbol (

) may also be displayed with error messages.

4.2 Troubleshooting

4.3 Transferring the Data to the Online 

Solution

To begin using the software for the first time, please 
visit www.omron-healthcare.com for the initial set-up 
instructions.

4.4 Maintenance

To protect your unit from damage, please observe the 
following:

• Do not subject the main unit and the cuff to 

extreme temperatures, humidity, moisture or 
direct sunlight.

• Do not disassemble the unit.
• Do not subject the unit to strong shocks or 

vibrations (for example, dropping the unit on the 
floor).

• Do not use volatile liquids to clean the main unit.
• Do not wash the cuff or immerse it in water.
• Do not use petrol, thinners or similar solvents to 

clean the cuff.

• Do not carry out repairs of any kind yourself. If a 

defect occurs, consult your OMRON retail outlet 
or distributor as mentioned on the packaging.

• The unit should be cleaned with a soft, dry cloth.
• Use a soft, moistened cloth and neutral soap to 

clean the cuff.

• Keep the unit in its storage case when not in 

use.

• Fold the cuff into the storage case.

Do not store the unit in the following situations:

• If the unit is wet.
• Locations exposed to extreme temperatures, 

humidity, direct sunlight, dust or corrosive 
vapours.

• Locations exposed to vibrations, shocks or 

where it will be at an angle.

Calibration and Service

• The accuracy of this blood pressure monitor has 

been carefully tested and is designed for a long 
service life.

• It is generally recommended to have the unit 

inspected every two years to ensure correct 
functioning and accuracy. Please consult your 
authorised OMRON dealer or the OMRON 
Customer Service at the address given on the 
packaging or attached literature. 

5. Technical Data

Note: Subject to technical modification without prior notice. 

• This device fulfils the provisions of EC directive 93/42/EEC 

(Medical Device Directive). 

• This blood pressure monitor is designed according to the European 

Standard EN1060, Non-invasive sphygmomanometers Part 1: 
General Requirements and Part 3: Supplementary requirements 
for electromechanical blood pressure measuring systems.

• This OMRON product is produced under the strict quality system of 

OMRON HEALTHCARE Co. Ltd., Japan. The Core component for 
OMRON blood pressure monitors, which is the Pressure Sensor, is 
produced in Japan.

Made in China

Alternating 
date/time 
display

This week 

Morning average

Note: If there are not enough 

measurements taken in a week 
to calculate the average the 
display will appear as 
illustrated.

This week 

Evening average

Last week 

Morning average

Last week 

Evening average

7 weeks ago 

Evening average

7 weeks ago 

Morning average

First measurement of 
the morning

Within 10 minutes

11:59

4:00

Measurements of the morning

Up to 3 measurements

Morning Blood Pressure

Measurement

Measurement

Measurement

Measurement

Measurement

Measurement

Sunday

Monday

Tuesday

Wednesday

Thursday

Friday

Saturday

Measurements of the evening

Last measurement of 
the evening

Within 10 minutes

1:59

19:00

Up to 3 measurements

Evening Blood Pressure

Measurement

Measurement

Measurement

Measurement

Measurement

Measurement

Sunday

Monday

Tuesday

Wednesday

Thursday

Friday

Saturday

Error Display

Cause

Remedy

Irregular heartbeats 
are detected.

Remove the wrist cuff. Wait 
2 - 3 minutes and then take 
another measurement.
Repeat the steps in section 
3.3. If this error continues to 
appear, contact your doctor.

Movement during 
measurement.

Carefully read and repeat the 
steps in section 3.3.

Blink

The batteries are 
low.

You should replace them with 
new ones ahead of time.
Refer to section 2.1.

Lit

The batteries are 
exhausted. 

You should replace them with 
new ones at once.
Refer to section 2.1.

Wrist cuff not 
applied correctly.

Apply the wrist cuff correctly.
Refer to section 3.1.

Movement during 
measurement.

Repeat Measurement while 
remaining still and refraining 
from talking during the 
measurement.
Refer to section 3.3.

Wrist cuff not 
applied correctly, or 
movement during 
measurement.

Apply the wrist cuff correctly 
and repeat measurement 
while remaining still and 
refraining from taking during 
the measurement.
Refer to section 3.1 and 3.3.

Arm position 
changed during 
measurement.

Remain still until the 
measurement is complete. 
Refer to section 3.3.

Device error.

Contact your OMRON retail 
outlet or distributor.

Note:It flashes to show 

which week 
average value is 
displayed.

Systolic blood 

pressure

above169 mmHg

152 - 168 mmHg

136 - 151 mmHg

119 - 135 mmHg

118 mmHg or less

3 weeks ago
7 weeks ago

2 weeks ago
6 weeks ago

Last week

5 weeks ago

This week

4 weeks ago

This week - 3 weeks ago

This week

morning SYS average

118 mmHg

This week

evening SYS average

140 mmHg

4 weeks ago - 7 weeks ago

7 weeks ago

morning SYS average

112 mmHg

7 weeks ago

evening SYS average

138 mmHg

First

Second

Problem

Cause

Remedy

The reading is 
extremely low 
(or high).

The wrist cuff is not 
at heart level.

Measure while in the 
correct posture.
Refer to section 3.2.

The cuff is not 
wrapped snugly 
around the wrist.

Wrap the cuff correctly.
Refer to section 3.1.

The arms and 
shoulders are 
tense.

Relax and try taking the 
measurement again.
Refer to section 3.3.

Movement or 
talking during 
measurement.

Remain still and do not 
talk during Measurement.
Refer to section 3.3.

Wrist cuff pressure 
does not rise.

Air is leaking from 
the wrist cuff.

Consult your OMRON 
retail outlet or distributor.

Wrist cuff deflates 
too soon.

The wrist cuff is 
loose.

Apply the cuff correctly so 
that it is firmly wrapped 
around the wrist. Refer to 
section 3.1.

The blood pressure is different 
each time. The reading is 
extremely low (or high).

Blood pressure readings 
constantly vary with time 
of day and how relaxed 
you are. Take several 
deep breaths and try to 
remain relaxed before 
taking a measurement.

The unit loses 
power during 
measurement.

The batteries are 
exhausted.

Replace the batteries with 
new ones.
Refer to section 2.1.

Nothing happens 
when you press 
the buttons.

The batteries have 
been inserted 
incorrectly.

Insert the batteries with 
the correct (+/-) polarity.
Refer to section 2.1.

Other problems.

• Press the START/STOP button and repeat 

measurement.

• If the problem continues, try replacing the 

batteries with new ones.

If this still does not solve the problem, contact 
your OMRON retail outlet or distributor.

Product Description

Wrist Blood Pressure Monitor

Model

OMRON RS8 (HEM-6310F-E)

Display

LCD Digital Display

Measurement Method

Oscillometric method

Transmission Method 

NFC-F (conforming to ISO/IEC)

Measurement Range

Pressure: 0 to 299 mmHg 
Pulse: 40 to 180 beats/min.

Accuracy

Pressure: ±3 mmHg 
Pulse: ±5% of display reading

Inflation

Automatic inflation by pump

Deflation

Automatic rapid deflation

Memory

100 Measurements with date and time for 
each user (A and B)

Power Source

2 x 1.5V (LR03, AAA alkaline batteries)

Battery Life

Approx. 300 measurements with new 
alkaline batteries at a room temperature of 
23°C

Applied Part

= Type B

Protection Against 
Electric Shock

Internally powered ME equipment

Operating temperature/ 
Humidity

+10 to +40°C / Maximum: 30 to 85% RH

Storage temperature/ 
Humidity/Air pressure

-20 to +60°C / Maximum: 10 to 95% RH / 
700 to 1060 hPa

Console Weight

Approximately 80 g without batteries

Outer Dimensions

Approximately 89 (w) mm x 61 (h) mm x 
13 (d) mm (without the wrist cuff)

Measurable 
circumference

Approximately 13.5 to 21.5 cm

Cuff Material

Nylon and polyester

Package Content

Main unit, storage case, battery set, 
instruction manual, guarantee card, blood 
pressure pass

Important information regarding Electro Magnetic 
Compatibility (EMC)

With the increased number of electronic devices such as PC’s and 
mobile (cellular) telephones, medical devices in use may be 
susceptible to electromagnetic interference from other devices. 
Electromagnetic interference may result in incorrect operation of 
the medical device and create a potentially unsafe situation.
Medical devices should also not interfere with other devices.

In order to regulate the requirements for EMC (Electro Magnetic 
Compatibility) with the aim to prevent unsafe product situations, the 
EN60601-1-2:2007 standard has been implemented. This standard 
defines the levels of immunity to electromagnetic interferences as 
well as maximum levels of electromagnetic emissions for medical 
devices.

This medical device manufactured by OMRON HEALTHCARE 
conforms to this EN60601-1-2:2007 standard for both immunity and 
emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which 

generate strong electrical or electromagnetic fields, near the 
medical device. This may result in incorrect operation of the unit 
and create a potentially unsafe situation. Recommendation is to 
keep a minimum distance of 7 m. Verify correct operation of the 
device in case the distance is shorter.

Further documentation in accordance with EN60601-1-2:2007 is 
available at OMRON HEALTHCARE EUROPE at the address 
mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.

Correct Disposal of This Product 
(Waste Electrical & Electronic Equipment)

This marking shown on the product or its 
literature, indicates that it should not be 
disposed of, with other household wastes at the 
end of its working life. 
To prevent possible harm to the environment or human health from 
uncontrolled waste disposal, please separate this product from 
other types of wastes and recycle it responsibly to promote the 
sustainable reuse of material resources.

Household users should contact either the retailer where they 
purchased this product, or their local government office, for details 
of where and how they can return this item for environmentally safe 
recycling.
Business users should contact their supplier and check the terms 
and conditions of the purchase contract. This product should not be 
mixed with other commercial wastes for disposal. 

Manufacturer

OMRON HEALTHCARE Co., Ltd.

53, Kunotsubo, Terado-cho, Muko, Kyoto, 
617-0002 JAPAN

EU-representative

OMRON HEALTHCARE EUROPE B.V.

Scorpius 33, 2132 LR Hoofddorp 
THE NETHERLANDS
www.omron-healthcare.com

Production Facility

OMRON (DALIAN) CO., LTD.

Dalian, CHINA

Subsidiary

OMRON HEALTHCARE UK LTD.

Opal Drive, Fox Milne, Milton Keynes, MK15 0DG 
U.K.

OMRON MEDIZINTECHNIK 
HANDELSGESELLSCHAFT mbH

John-Deere-Str. 81a 68163 Mannheim, GERMANY
www.omron-medizintechnik.de

OMRON SANTÉ FRANCE SAS

14, rue de Lisbonne, 93561 Rosny-sous-Bois 
Cedex, FRANCE

HEM-6310F-E.book  Page 2  Monday, December 10, 2012  4:14 PM

Отзывы: