Troubleshooting

Issue

Inspection location

Solution

The power LED 
does not light up.

Is the power cord connected correctly to a power 
outlet and the main unit?

Plug the product into an outlet.  

 Page 9)

Nebulization is not 
functioning.

Is the “

 WATER LEVEL” Error LED lit up due to 

there being too little water in the removable water tank?

Fill the tank with water to the optimal water 
level.  

 Page 7)

Have the medication cup and removable water tank 
been attached properly?

Attach the parts properly.  

 Page 8)

Is the fan cover open?

Close the cover tightly.  

Has the fan been attached?

Attach the fan properly.  

 Page 18)

Is there too much medication in the medication cup?

Reduce the amount of medication to 150 ml or 
less.  

 Page 8)

Did you add distilled or purified water to the 
removable water tank?

Add tap water to the tank.  

Are the contacts on the main unit or removable water 
tank contaminated?

Wipe away any contamination with a cloth that 
has been moistened with water or sanitizing 
alcohol and then wrung out lightly.  

Is the inhalation hose bent, causing medication to 
accumulate inside it?

Straighten the inhalation hose and remove any 
accumulated medication.  

Has the START/STOP button been turned on?

Press the START/STOP button. (   Page 11)

The aerosol 
volume is low.
The nebulization 
capacity is low.

Is there too little medication in the medication cup?

Add more medication so that there is at least 
10 ml in the cup.  

 Page 8)

Is the medication cup damaged or misshapen? 

Replace the medication cup with a new one.

Is there too much medication in the medication cup?

Reduce the amount of medication so that it is 
150 ml or less.  

 Page 8)

Is the nebulization volume or air-flow volume setting 
too low?

Increase the setting.  

 Page 10)

Is the room temperature or water temperature too 
low?

Replace the water in the removable water tank 
with water whose temperature is about 26 °C. 

Is the medication cup contaminated?

Disinfect the medication cup. If unable to 
remove the contamination, replace the 
medication cup with a new one.  
 ( 

 Page 16 and Page 20)

Nebulization is 
inconsistent.

Is there too much medication in the medication cup?

Reduce the amount of medication so that it is 
150 ml or less.  

 Page 8)

Does the medication cup contain 150 ml of 
medication, and are the maximum values for both the 
air-flow volume and nebulization volume settings 
being used?

Adjust the nebulization volume.  (   Page 10)

Issue

Inspection location

Solution

Error LEDs

LED 

 is lit up.

There is too little water in the 
removable water tank.

Fill the tank with water to the optimal 
water level.  

 Page 7)

LED 

 is lit up.

The medication cup cover is open.

Close the cover tightly.  (   Page 8)

LED 

 is lit up.

The fan cover is open.

Close the cover tightly.

LED 

 is flashing.

It’s time to inspect the air filter.
(The LED lights up after about 100 
hours of nebulization.)

Inspect the air filter. If it is excessively 
contaminated, replace it with a new air filter. 
  

 Page 17)

E1 indicator

The main unit’s internal fan has 
stopped.

Unplug the cord, label the main unit to 
indicate that it is broken, and contact an 
authorised OMRON retail outlet or 
distributor.

E3 indicator

The main unit’s internal software 
experienced an error.

E4 indicator

There is a short in the main unit’s 
internal oscillation circuit. 

E5 indicator

The removable water tank has not 
been secured properly.

Remove the removable water tank and 
then secure it in the main unit.  

 

 ( 

 Page 9)

Note:  If the suggested remedy does not solve the problem, do not try to repair the device - the device is not user serviceable. 

Return the device to an authorised OMRON retail outlet or distributor.

Содержание NE-U780

Страница 1: ...throat EN FR DE IT ES NL RU AR Table of Contents Important Information Intended Use 1 Important Safety Precautions 2 Exemptions 3 Important Information Know Your Unit Product Features 4 Packaging Cont...

Страница 2: ...of a medical expert Some hospitals clinics may rent out the devices Durable period Durable periods are as follows provided the product is used to nebulize 30 ml of saline solution at 26 C with 26 C w...

Страница 3: ...ain unit contact are not contaminated Make sure that the air filter is correctly attached To avoid the medication residue on the face be sure to wipe the face after removing the mask Do not get nebuli...

Страница 4: ...failure or damage or accidents involving the product resulting from a failure to adhere to the safety precautions and operating instructions described in this manual 5 Product failure or damage result...

Страница 5: ...E ZKLFK WKH medication becomes an aerosol 1HEXOL DWLRQ 7KH SURFHVV E ZKLFK WKH DWRPL HG medication is expelled by the product 1 Smooth shape for easy cleaning The main unit is easy to clean and disin...

Страница 6: ...hen using the timer Indicates the remaining nebulization time The time can be set from 1 to 30 min During continuous nebulization The digital portion of the display cycles through a simple animation P...

Страница 7: ...f main unit Top of main unit Fuse Electric power socket Power cord Display Handle Mouthpiece Inhalation hose Fan cover Fan Air filter case Air filter Vibrator Nebulization set Medication cup cover Sil...

Страница 8: ...acing your finger under the protrusion on the medication cup holder remove the nebulization set Protrusion 3 Fill the removable water tank with tap water Fill to the optimal water level taking care no...

Страница 9: ...to the medication cup Min 10 ml Max 150 ml Notes HIRUH DGGLQJ PHGLFDWLRQ PDNH VXUH WKH medication cup is clean not damaged or misshapen I XVLQJ D V ULQJH WDNH FDUH QRW WR GDPDJH WKH medication cup wit...

Страница 10: ...power cord 4 Turn on the power button and make sure all the Error LEDs on the error display light up and then turn back off If one or more of the Error LEDs remain lit see Troubleshooting Page 22 for...

Страница 11: ...ULIZATION VOLUME adjustment buttons set the air flow volume and nebulization volume You can check both settings with the level displays which change to reflect the current value of each The air flow v...

Страница 12: ...ion mask optional can be used instead of the mouthpiece During inhalation Do not bend the inhalation hose or block the mouthpiece It may cause medication to accumulate in the hose reducing the nebuliz...

Страница 13: ...mple display with the timer set to 10 min When operation stops After about 5 sec This prepares the time for the next patient To stop inhalation while using the timer Press the START STOP button to sto...

Страница 14: ...fan cover for bacterial filter Instead of the normal fan cover use the optional NEB BFC 78E Fan Cover for Bacterial Filter 2 Remove the air volume adjustment lever on the long time nebulization set an...

Страница 15: ...an be connected to the product Continuous nebulization of medication volumes in excess of 150 ml can be accomplished by using the optional NEB LMCC 78E Long time Nebulization set and U10 11 E Supply M...

Страница 16: ...is not bent Medication will not be supplied if the tube is bent If the tube is too long cut it to a suitable length For subsequent steps see No 2 under Step 3 in the Before Inhalation section Page 9 S...

Страница 17: ...mouthpiece and other accessories from the main unit 3 Remove the removable water tank from the main unit 4 Remove the nebulization set from the removable water tank and disassemble it 5 Dispose of an...

Страница 18: ...e air filter is not contaminated If the air filter is excessively contaminated replace it with a new air filter 4 5 Return the air filter case and fan cover to their original positions and align the m...

Страница 19: ...t clean 5 Reattach the fan air filter case and fan cover When reattaching the fan push it firmly all the way onto the fan installation axis 6 Align the medication cup cover so that it is once more con...

Страница 20: ...infectant see its instruction manual and precautionary information Some disinfectants may trigger an allergic reaction Although the amount of time for which parts should be immersed varies with the di...

Страница 21: ...diene styrene resin ABS Poly phenylene sulfide resin PPS Polyethylene resin PE Polycarbonate resin PC Polymethylpentene PMP Polypropylene resin PP Polystyrene resin PS Polyvinyl chloride resin PVC The...

Страница 22: ...icon rubber Silicon packing Washer Silicon rubber Bottle PP Silicon packing Washer Silicon rubber Tube Silicon rubber Hanger Stainless steel Long time Nebulization set Supply Medication set Do not dis...

Страница 23: ...the nebulization volume or air flow volume setting too low Increase the setting Page 10 Is the room temperature or water temperature too low Replace the water in the removable water tank with water w...

Страница 24: ...tank Medication cup cover Medication cup holder Inhalation hose M Mouthpiece Medication cup 2 pcs Power cord Instruction Manual including product warranty Classifications Class I Protection against e...

Страница 25: ...th the aim to prevent unsafe product situations the EN60601 1 2 2007 standard has been implemented This standard defines the levels of immunity to electromagnetic interferences as well as maximum leve...

Страница 26: ...aminants out of the airflow Bacterial Filter Materials PP PMP Model U17 2 E Remarks Includes adapter Used for inhalation when cleansing the respiratory tract etc Cannot be reused Fan cover for bacteri...

Страница 27: ...rt b Costs for repairs and or defects resulting from repairs done by unauthorised persons c Periodic check ups and maintenance d Failure or wear of accessories or other attachments other than the main...

Страница 28: ...EALTHCARE Co Ltd Matsusaka Factory 1855 370 Kubo cho Matsusaka shi Mie 515 8503 Japan Subsidiaries OMRON HEALTHCARE UK LTD Opal Drive Fox Milne Milton Keynes MK15 0DG UK www omron healthcare com OMRON...

Отзывы: