![Omron MicroAIR U22 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/omron/microair-u22/microair-u22_instruction-manual_743925019.webp)
19
The mesh cap is a consumable part.
/MRON RECOMMENDS TO EXCHANGE THE MESH CAP AFTER APPROXIMATELY YEAR
Remove the mesh cap and install a new one as described on page 10 (section “How to assemble the
unit and fill the medication container”).
!DDITIONAL INSTRUCTIONS FOR -ESH CARE
In case the performance of the unit remains reduced even after cleaning and disinfecting, use the
following methods to remove dried medication, lime or saliva which may clog the tiny holes in the mesh
cap.
Leave the mesh cap in a white vinegar solution (1 part vinegar diluted with 3 parts of water) for
½ - 1 hour. In addition, the mesh cap may be kept overnight in a cup of distilled water, in which a tablet
of dental (“false teeth”) cleaner has been added. After cleaning with vinegar or dental cleaner, be sure
TO SHAKE OFF EXCESSIVE SOLUTION THOROUGHLY BUT CAREFULLY RINSE THE MESH WITH DISTILLED WATER 3HAKE OFF
EXCESSIVE WATER FROM THE MESH CAP AND LET IT AIRDRY ON A CLEAN LINT FREE PAPER TOWEL
In case the unit nebulizes poorly even after following this procedure, replace the mesh cap.
(OW TO REPLACE THE MESH CAP
Содержание MicroAIR U22
Страница 1: ...Instruction manual s NGLISH s EUTSCH s EDERLANDS s U22 IM NE U22 E 08 01 2014 1665580 0F...
Страница 27: ......