17

How to Disinfect the Unit

How to Replace the Mesh Cap

2

Remove the Mesh Cap
from the Medication Bottle

3

Install a new Mesh Cap.

- The nebulization rate may vary

depending on the type of Mesh 
Cap you use.

You can usually use a Mesh Cap for approximately 1 year (according to the test result of
OMRON).  However, the unit may not nebulize even before using it a year depending on
how you use it.  The unit also may not nebulize or nebulize extremely weak even after
disinfected by boiling.  In those cases, replace the Mesh Cap with a new one. 

(Mesh Cap is a consumable part.) 

For the inquiry on where to purchase the Mesh Cap or the broken or lost accessory,
contact the store where you purchased the unit or the nearest OMRON dealer.

Medication Bottle
Disengaging Button

1

Remove the Medication
Bottle from the Main Unit.

- Lift up the Medication Bottle from

the Main Unit by pushing in the
Medication Bottle Disengaging
Buttons from both sides and
releasing the engagement.

Lift up the Mesh Cap by
applying your thumb on
the brim.

Lift up the
Mesh Cap
perpendicular
to the
Medication
Bottle.

Do not insert your 
finger into the 
Nebulization Port.

Do not insert your 
finger into the 
Nebulization Port.

Insert the Mesh Cap 
perpendicular to the 
Medication Bottle.

Align these points,
then push in as far
as it goes.

Close the Mesh Cap

securely.

How to Replace the Mash Cap

EN-p2-20-(NE-U22)-3  3/25/02  4:48 PM  Page 17

Содержание Micro A-I-R NE-U22

Страница 1: ...cifications Option parts 22 This unit is a medical instrument Be sure to follow the instructions of a doctor and use the unit correctly For the type dose and regimen of medication be sure to follow th...

Страница 2: ...h wet hands You may suffer electric shock or get burned Do not wash the Main Unit and AC Adapter with water or hot water Do not immerse them in water The unit may experience electricity leakage or you...

Страница 3: ...when the unit is tilted To start inhalation you must assemble the unit Please read the section on How to Assemble the Main Unit and understand the points for assembling before starting to use the uni...

Страница 4: ...ator vibrates at high frequency and pushes out medication from the pores of Mesh Electrode A contact point connecting the Main Unit and the Medication Bottle Battery Low Indicator An orange light flas...

Страница 5: ...ap Locking Lever Opens the Medication Bottle Cap Do not open Medication Bottle Cap except when washing Medication Bottle Medication Bottle Cap Medication Bottle Stores the medication Medication Bottle...

Страница 6: ...stallation is complete Installation is complete Do not open the Bottle Cap Do not insert your finger into the Nebulization Port Insert the Mesh Cap perpendicular to the Medication Bottle Align these p...

Страница 7: ...rage Bag or in a clean place How to install the Inhalation Mask S How to install the Mouthpiece Installation is complete Installation is complete Installation is complete Securely cover the Main Unit...

Страница 8: ...c outlet 2 Unlock the Main Unit by pushing in both ends of the Main Unit Locking Board of the AC Adapter Connection Stand with your fingers 2 Plug in the power plug of AC Adapter to the electric outle...

Страница 9: ...ries with new ones When the Battery Low Indicator orange turns on batteries are worn out As the unit will not nebulize immediately replace both batteries with new ones You may use manganese batteries...

Страница 10: ...Bottle is 7 ml 4 Install the Mesh Cap If the Mesh Cap is not securely closed medication may leak Medication Bottle Disengaging Button Insert the Mesh Cap perpendicular to the Medication Bottle Do not...

Страница 11: ...e off the stain 6 Install the Mask Adapter 7 Install the Inhalation Mask or the Mouthpiece Align both Electrodes with each other Keep the Electrodes clean How to install the Inhalation Mask S How to i...

Страница 12: ...eep pressing the Button approximately for 2 seconds or more the unit nebulizes in Manual Nebulization Mode The unit nebulizes only while you keep on pressing the Button and stops nebulization when you...

Страница 13: ...r in the medication If the Vibrator is not immersed such as when the Main Unit is placed on a horizontal surface the unit may not nebulize Medication Vibrator Air Port If the medication has high surfa...

Страница 14: ...ity or viscosity such as medication solubilizing agent or expectorant because the medication is often left on the Mesh Please be careful not to spray water on the Main Unit Be sure to clean and store...

Страница 15: ...n Port General advice Never try to wipe off water with tissues or cloth Dust or lint left on the Mesh may affect nebulization General advice Do not wipe the Main Unit with volatile agent such as benze...

Страница 16: ...must be done after confirming the target bacteria and the effects of disinfection method Do not disinfect the unit with the disinfection solution containing sodium hypochlorite The unit may not be abl...

Страница 17: ...chase the Mesh Cap or the broken or lost accessory contact the store where you purchased the unit or the nearest OMRON dealer Medication Bottle Disengaging Button 1 Remove the Medication Bottle from t...

Страница 18: ...torage bag to avoid the Main Unit being scratched For Optional AC adapter Bundle the power cord of AC Adapter and bind it with the AC Adapter using a Cord Band as shown in the Figure General advice Do...

Страница 19: ...utton towards yourself Refer to Page 13 Remove the medication or water Refer to Page 13 Wipe off the stain Refer to Page 15 If the unit does not nebulize normally after taking the above mentioned proc...

Страница 20: ...out 30 minutes a day with the use of alkaline batteries Operating temperature and humidity 10 to 40 C 30 to 85 RH Accessories Inhalation Mask S Inhalation Mask L Mouthpiece Mask Adapter Storage Bag In...

Отзывы: