20
Within Run Precision:
Samples
Mean glucose concentration
measured CV
Sample 1
41.01 mg/dL
83.49 mg/dL
116.82mg/dL
177.52mg/dL
177.52mg/dL (9.85 mmol/L)
341.61mg/dL
177.52mg/dL (9.85 mmol/L)
42.83 mg/dL
134.36 mg/dL
334.93 mg/dL
Sample 2
4.26%
Sample 3
4.85%
Sample 4
4.66%
Sample 5
4.05%
Between Day Precision:
Control solution 1
Control solution 2
4.20%
Control solution 3
3.44%
Chapter 8 Specifications
8.1 Specification of HGM-112 meter
◆
Measuring range: 20~600mg/dL
◆
Calibration: Plasma-calibration
◆
Measuring unit: mg/dL
◆
Measuring time: 5 seconds
◆
Sample type: Capillary whole blood samples drawn from the fingertips
or palm
◆
Blood Sample size: 1μL
◆
Operating temperature: (10 C ~ 40 C) / (50 ºF ~ 104ºF)
◆
Relative operating humidity: Below 90%RH
◆
Memory capacity: one memory for the lastest memory
◆
Dimension: 63.2 mm x 40.6 mm x 18.0 mm (L×W×H)
◆
Weight: 25.5 g with battery
◆
Power Supply: one CR2032 battery
◆
Storage temperature: -20 C ~ 50 C (-4 F ~ 122 F)
◆
Please protect the HGM-112 from impact
◆
Please do not disassemble the HGM-112
SD= 4.02 mg/dL
3.61 mg/dL
Precision:
This study shows the variability from strip to strip in sample tests. The
results are shown in the following table.
BLOOD GLUCOSE MONITORING SYSTEM-HGM-112