39
Recommended separation distance between portable and mobile RF communications
equipment and the OMRON BPM including AC-adapter
OMRON BPM including AC-adapter is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of this OMRON BPM
including AC-adapter can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the OMRON
BPM including AC-adapter as recommended below, according to the maximum output power of the
communications equipment.
Output Power of Transmitter in Watt
Separation distance according to frequency of
transmitter in meter
150 kHz to
80 MHz
d
= 1.2
80 MHz to 800 MHz
d
= 1.2
800 MHz to 2.5GHz
d
= 2.3
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
P
P
P