23
Air plug disconnected.
Insert the plug securely.
Refer to section 3.1.
Arm cuff is applied too
loosely.
Apply the arm cuff tighter.
Refer to section 3.1.
Air is leaking from the
arm cuff.
Replace the arm cuff with a
new one.
Refer to section 5.3.
Movement during
measurement and the arm
cuff has not been inflated
sufficiently.
Repeat measurement. Remain
still and do not talk during
measurement.
Refer to section 3.3.
If “E2” appears repeatedly, inflate
the arm cuff manually until it is 30
to 40 mmHg above your previous
measurement result.
Refer to section 3.3.
The arm cuff was inflated
exceeding the maximum
allowable pressure, and then
deflated automatically when
inflating the arm cuff
manually.
Do not touch the arm cuff and/or
bend the air tube while taking a
measurement. Do not inflate the
arm cuff more than necessary.
Refer to section 3.3.
Error Display
Cause
Solution