56
Display
Cause
Solution
Please see
the manual.
Error
7
Your wrist is moved
up and down during
a measurement.
Take another measurement making sure not
to move your wrist. Refer to sub-section 3.1
(page 25).
Please see
the manual.
Error
8
Temperature error is
detected.
Wait for few minutes, then take another
measurement in a place with moderate room
temperature. If the error appears again, contact
customer service. Refer to section 13
(page 69).
The monitor has
malfunctioned.
Please see
the manual.
Error
9
The band may be
applied too tightly.
Apply the band correctly, then take another
measurement. Refer to sub-section 2.2
(page 21).
Error
Room temperature
is too high to charge
the battery.
Charge the battery at a moderate room
temperature.
The monitor has
malfunctioned.
If the error appears in moderate temperature,
contact customer service. Refer to section 13
(page 69).
8. Error Messages and Troubleshooting