26
Sample 2
99.6 mg/dL(5.5mmol/L)
3.0%
Sample 3
152.4 mg/dL(8.5mmol/L)
2.8%
Sample 4
247.8 mg/dL(13.8mmol/L)
2.9%
Sample 5
400.0 mg/dL(22.2mmol/L)
3.1%
Between Day Precision
Control solution 1 56.2 mg/dL(3.1mmol/L)
SD=4.2 mg/dL (0.2
mmol/L)
Control solution 2 104.1 mg/dL(5.8mmol/L)
3.2%
Control solution 3 306.1 mg/dL(17.0mmol/L)
2.9%
Chapter 9 Specification and Maintenance
◆
Measuring range: 20~600mg/dL or 1.1~33.3mmol/L
◆
Measuring unit: mg/dL or mmol/L
◆
Measuring time: Approximately 5 seconds
◆
Blood Sample size: Approximately 1μL
◆
Operating temperature: (10
℃
~40
℃
)/(50 ºF ~104ºF)
◆
Store the test strips at ambient temperature between
4~30
℃
(39ºF ~86ºF). Do not refrigerate.
◆
Relative operating humidity: Below 90% RH
◆
Relative storage humidity: Below 75% RH
◆
Memory capacity: Up to 512 test results (together with
date, time and unit) include average results for the last 7,
14 and 28 days
◆
Dimension: 9.0 x 6.0 x 2.5 cm (L×W×H)
◆
Weight: Approximately 90g with batteries
◆
Power supply: Two 1.5V AAA batteries
◆
Recommend to store in room temperature environment.
◆
Please do not store in high temperature or high humidity.
◆
Please prevent the HEA-230 system against from impact.
◆
Please do not disassemble tthe HEA-230 system.
warning: The measurement unit either mg/dL or mmol/L is fixed in the
meter according to the country demand. You are not able to change the
unit in the meter.
Caution: Any further service and product information, please contact
OMRON distributors.
OMRON blood glucose monitoring system HEA-230.indd 26
2012/6/14 上午 11:14:56
Содержание HEA-230
Страница 2: ...OMRON blood glucose monitoring system HEA 230 indd 3 2012 6 14 上午 11 14 48 ...
Страница 30: ...OMRON blood glucose monitoring system HEA 230 indd 28 2012 6 14 上午 11 14 56 ...
Страница 31: ...OMRON blood glucose monitoring system HEA 230 indd 29 2012 6 14 上午 11 14 56 ...
Страница 32: ...5011554500 12 06 REV 00 OMRON blood glucose monitoring system HEA 230 indd 30 2012 6 14 上午 11 14 56 ...