TROUBLESHOOTING GUIDE
12
PROBLEM
CAUSES AND SOLUTIONS
The device is on
but nebulizes
weakly.
• Make sure the nebulizer is seated or check for proper
air tubing connection; repeat the procedure of insertion
as described in the manual.
• Make sure the nebulizer is in an upright position
during use. Do not tilt beyond 30º.
• Check whether the nebulizer is empty.
• Clean or replace the nebulizer as described in the
CLEANING AND MAINTENANCE section.
• Check whether the air filter is dirty or obstructed. If
necessary, replace it as described in the CLEANING
AND MAINTENANCE section.
The device is on
but does not
nebulize.
• Make sure the micronization blue cone has been placed
inside the nebulizer chamber.
The device does
not work and
the light indica-
tor is red.
The thermal protector has been activated for the following
reasons:
• the device has been operating beyond the operating
conditions indicated in this manual,
• the device has been operating near heat sources or in
hot environments.
Let the unit cool down for at least 40 minutes, wait until
the red light turns off and turn the unit on again.
Should the device not start working properly again,
refer to an authorized technical service center.