2 Installation and connection
2-4 Connecting the AC input
BU5002R/BU3002R/BU3002RH
55
2
3.
Loosen the screw on the terminal block using a screwdriver, insert the wire, and
tighten the screw on the terminal block.
The wire to be connected to the G terminal should be longer than the wires to be
connected to L1 and L2.
See Table 1 for the wire size to be used.
<BU5002R>
<BU3002RH>
When installing the G terminal of the
BU5002R, pass it below the L1 and
L2 wires, and arrange the wiring
horizontally.
G terminal: Connect the ground
wire
L1 terminal: Connect the line
L2 terminal: Connect the neutral line
Table 1
BU5002R
BU3002RH
Round terminal size
O.D.: Less than 12.7mm
I.D.: M5
O.D.: Less than 9.6 mm
I.D.: M4
Recommended wiring
outside diameter of clamp
9 to 15 mm
7 to 12 mm
Tightening torque
2.0 N·m
1.2 N·m
Recommended cable size
3.5 mm
2
(AWG 12)
2 mm
2
(AWG 14)