13. Guidance and Manufacturer’s Declaration
EN
Guidance and manufacturer’s declaration - electromagnetic immunity
OMRON BPM including AC-adapter is intended for use in the electromagnetic environment specified below. The customer or the user of this
OMRON BPM including AC-adapter should assure that it is used in such environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floor should be wood, concrete, or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply lines
±1 kV for input/output lines
Mains power quality should be that of a
typical commercial and/or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Mains power quality should be that of a
typical commercial and/or hospital
environment.
Voltage dips, short interruptions
and voltage variations on power
supply inputlines
IEC 61000-4-11
<5 %
U
T
(>95 % dip in
U
T
)
for 0.5 cycle
<5 %
U
T
(>95 % dip in
U
T
)
for 0.5 cycle
Mains power quality should be that of a
typical commercial and/or hospital
environment. If the user of the OMRON BPM
including AC-adapter requires continued
operation during power mains interruption, it
is recommended that the OMRON BPM
including AC-adapter be powered from an
uninterruptible power supply.
40 %
U
T
(60 % dip in
U
T
)
for 5 cycles
40 %
U
T
(60 % dip in
U
T
)
for 5 cycles
70 %
U
T
(30 % dip in
U
T
)
for 25 cycles
70 %
U
T
(30 % dip in
U
T
)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T
)
for 5 sec.
<5 %
U
T
(>95 % dip in
U
T
)
for 5 sec.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital environment.
Note:
U
T
is the A.C. mains voltage prior to application of the test level.