9. Guidance and Manufacturer’s Declaration

Guidance and manufacturer’s declaration - electromagnetic immunity

OMRON BP765CAN including AC-adapter is intended for use in the electromagnetic environment specified below. The customer or the user of 

this OMRON BP765CAN including AC-adapter should assure that it is used in such environment.

Immunity test

IEC 60601 test level

Compliance 

level

Electromagnetic environment - guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 V rms

150 kHz to 80 MHz

3 V/m

80 MHz to 2.5 GHz

3 V rms

3 V/m

Portable and mobile RF communications equipment should be used no closer to 

any part of the OMRON BP765CAN including AC-adapter including cables, than 

the recommended separation distance calculated from the equation appropriate to 

the frequency of the transmitter.

Recommend separation distance

d = 1.2 

P

d = 1.2 

P

80 MHz to 800 MHz

d = 2.3 

P

800 MHz to 2.5 GHz

where P is the maximum output power rating of the transmitter in watts (W) 

according to the transmitter manufacturer and d is the recommended separation 

distance in meters (m).

Field strengths from fixed RF transmitters as determined by an electromagnetic 
site survey,

a

 should be less than the compliance level in each frequency range.

b

Interference may occur in the vicinity of equipment marked with the following 

symbol:

Note1: At 80 MHz and 800 MHz, the higher frequency range applies.

Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, 

objects, and people.

a

 Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones and land mobile radio, AM and FM radio 

broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF 

transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the OMRON 

BP765CAN including AC-adapter is used exceeds the applicable RF compliance level above, the OMRON BP765CAN including AC-adapter 

should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as 

reorienting or relocating the OMRON BP765CAN including AC-adapter.

b

 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

HEM-7311-CACS_C_M.book  Page 32  Monday, April 21, 2014  12:10 PM

Содержание BP765CAN

Страница 1: ...a 2013 OMRON HEALTHCARE INC Made in China Fran ais 2298807 1C Fabriqu pour OMRON HEALTHCARE Co Ltd 53 Kunotsubo Terado cho Muko Kyoto 617 0002 JAPON Distribu par OMRON HEALTHCARE INC 1925 West Field C...

Страница 2: ...a 2013 OMRON HEALTHCARE INC Made in China Fran ais 2298807 1C Fabriqu pour OMRON HEALTHCARE Co Ltd 53 Kunotsubo Terado cho Muko Kyoto 617 0002 JAPON Distribu par OMRON HEALTHCARE INC 1925 West Field C...

Страница 3: ...ation The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with the measurement result Important Safety Information 1 1 Know Your Device 4 1 1 Displa...

Страница 4: ...evice on the injured arm or the arm under medical treatment Do not apply the arm cuff on the arm while being on an intravenous drip or blood transfusion Consult your physician before using the device...

Страница 5: ...eneral Precautions Do not forcibly crease the arm cuff or the air tube excessively Do not press the air tube while taking a measurement To unplug the air plug pull on the air plug at the connection wi...

Страница 6: ...er than measuring blood pressure Use only the approved arm cuff for this device Use of other arm cuffs may result in incorrect measurement results Do not use a mobile phone or other devices that emit...

Страница 7: ...USER ID symbol TruReadTM symbol Systolic blood pressure Diastolic blood pressure Heartbeat symbol Flashes during measurement Pulse display Memory number Average value symbol Date Time display Movement...

Страница 8: ...isplays with your measurement results we recommend you consult your physician Follow the directions of your physician Movement Error Symbol The movement error symbol is displayed if you move your body...

Страница 9: ...minutes before taking a measurement Rest for at least 15 minutes before taking the measurement 2 Stress raises blood pressure Avoid taking measurements during stressful times 3 Measurements should be...

Страница 10: ...ay turn the monitor off then replace all batteries at the same time Long life alkaline batteries are recommended The measurement values continue to be stored in memory even after the batteries are rep...

Страница 11: ...NOTES Use only the authorized OMRON AC Adapter that came with this monitor We recommend you install the batteries even if the AC Adapter is used If no batteries are installed you may need to reset th...

Страница 12: ...the right side of the monitor 2 Plug the AC adapter into an electrical outlet To disconnect the AC adapter unplug the AC adapter from the electrical outlet first and then remove the AC adapter plug f...

Страница 13: ...tton The TruReadTM symbol appears on the display 3 To set the date and time proceed to section 2 4 If the date and time are correct press the START STOP button to turn the monitor off 1 3 2 4 Press th...

Страница 14: ...nth day hour and minute 3 Press the START STOP button to turn the monitor off Notes If the batteries have been replaced the date and time setting will need to be reset If the date and time are not set...

Страница 15: ...er Notes When you take a measurement on the right arm the air tube will be at the side of your elbow Be careful not to rest your arm on the air tube The blood pressure can differ between the right arm...

Страница 16: ...g drinking alcohol or caffeine smoking exercising or eating 30 minutes before taking a measurement Sit on a chair with your feet flat on the floor Sit upright with your back straight Sit with your bac...

Страница 17: ...ect your USER ID 1 or 2 2 Press the START STOP button The arm cuff starts to inflate automatically If your systolic pressure is more than 210 mmHg After the arm cuff starts to inflate press and hold t...

Страница 18: ...tically turn off after 2 minutes Note Wait 2 3 minutes before taking another measurement Waiting between measurements allows the arteries to return to the condition prior to taking a measurement Using...

Страница 19: ...ay The second measurement is complete waiting to start the next measurement The interval will appear on the display The third measurement is complete the average for the 3 measurements appears on the...

Страница 20: ...se the START STOP button when the Date Time display turns off The arm cuff will start to inflate automatically DO NOT adjust medication based on measurement results from this blood pressure monitor Ta...

Страница 21: ...2 values If there is 1 measurement value in the memory for that period this is displayed as the average If the memory is full the monitor will delete the oldest value When viewing the measurement val...

Страница 22: ...dividual measurement values press the button while the average is displayed To View the Average Value 1 Select your USER ID 1 or 2 2 Press and hold the button for more than 3 seconds Notes If the prev...

Страница 23: ...USER ID 1 or 2 2 Press the Memory button while the memory symbol appears 3 While holding the button down press the START STOP button for more than 3 seconds Note You cannot partially delete the values...

Страница 24: ...tion 3 3 If this error continues to appear contact your physician Movement during measurement Carefully read and repeat the steps in section 3 3 The batteries are low You should replace the batteries...

Страница 25: ...t is 30 to 40 mmHg above your previous measurement result Refer to section 3 3 The arm cuff was inflated exceeding the maximum allowable pressure and then deflated automatically when inflating the arm...

Страница 26: ...k the battery installation for proper placement of the battery polarities Refer to section 2 1 Measurement values appear too high or too low Blood pressure varies constantly Many factors including str...

Страница 27: ...onents or immerse them in water Do not use gasoline thinners or similar solvents to clean the device Use a soft and dry cloth or a soft and moistened cloth and neutral soap to clean on the monitor and...

Страница 28: ...Put the arm cuff into the cuff compartment and close the monitor cover Do not store the device in the following situations If the device is wet Locations exposed to extreme temperatures humidity direc...

Страница 29: ...Maintenance and Storage 5 3 Optional Medical Accessories Arm Cuff Arm circumference 9 17 22 42 cm AC Adapter CD WR17 Model HEM RML31 HEM ADPTW5 HEM 7311 CACS_C_M book Page 26 Monday April 21 2014 12 1...

Страница 30: ...V 50 60Hz 0 12A Battery life Approximately 1000 measurements using new alkaline batteries Operating temperature humidity 50 F to 104 F 10 C to 40 C 15 to 90 RH Storage temperature humidity air pressur...

Страница 31: ...with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause ha...

Страница 32: ...receipt Include a letter with your name address phone number and description of the specific problem Pack the product carefully to prevent damage in transit Because of possible loss in transit we reco...

Страница 33: ...e observed to verify normal operation in the configuration in which it will be used Refer to further guidance below regarding the EMC environment in which the device should be used The MEDICAL ELECTRI...

Страница 34: ...ine s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial and or hospital environment Voltage dips short interruptions and voltage variations...

Страница 35: ...itters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the followin...

Страница 36: ...Power of Transmitter in Watt Separation distance according to frequency of transmitter in meter 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0...

Страница 37: ...ation The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with the measurement result Important Safety Information 1 1 Know Your Device 4 1 1 Displa...

Страница 38: ...ation The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with the measurement result Important Safety Information 1 1 Know Your Device 4 1 1 Displa...

Страница 39: ...a 2013 OMRON HEALTHCARE INC Made in China Fran ais 2298807 1C Fabriqu pour OMRON HEALTHCARE Co Ltd 53 Kunotsubo Terado cho Muko Kyoto 617 0002 JAPON Distribu par OMRON HEALTHCARE INC 1925 West Field C...

Отзывы: