Specifications
Notes:
• These specifications are subject to change without notice.
• In the clinical validation study, the 5th phase was used on 85 subjects for determination of diastolic blood pressure.
• This device has not been validated for use in pregnancy.
Model
BP765CAN HEM-7311-CACS
Display
LCD digital display
Measurement range
Pressure: 0 to 299 mmHg
Pulse: 40 to 180 beats / min.
Accuracy
Pressure:
±
3 mmHg or 2% of reading
Pulse:
±
5% of display reading
Inflation
Fuzzy-logic controlled by electric pump
Deflation
Automatic pressure release valve
Measurement method
Oscillometric method
IP classification
IP 20
Power source
4 “AA” batteries 1.5V or AC adapter (INPUT AC100-240V 50/60Hz 0.12A)
Battery life
Approximately 1000 measurements (using new alkaline batteries)
Operating temperature / humidity
50°F to 104°F (10°C to 40°C) / 15 to 90% RH
Storage temperature / humidity / air pressure
-4°F to 140°F (-20°C to 60°C) / 10 to 95% RH / 700 to 1060 hPa
Weight
Monitor : Approximately 1 lbs 6 5/8 oz. (640 g) not including batteries
Arm cuff : Approximately 6 oz. (170 g)
Dimensions
Monitor : Approximately 7 1/4” (w)
×
3 7/8” (h)
×
9” (l)
(183 mm
×
99 mm
×
230 mm)
Arm cuff : Approximately 5 3/4”
×
23 3/8” (air tube: 29 1/2”)
(145 mm
×
594 mm (air tube: 750 mm))
Cuff circumference
9” to 17” (220 to 420 mm)
Memory
Up to 100 per user
Contents
Monitor, arm cuff, AC adapter, instruction manual, quick start guide
Applied part
Protection against electric shock
Internally powered ME equipment (When using only the batteries)
= Type BF
= Class II ME equipment (AC adapter)
HEM-7311-CACS_C_M.book Page 27 Monday, April 21, 2014 12:10 PM