27
EN
6.
Specifications
Notes:
• These specifications are subject to change without notice.
• In the clinical validation study, the 5th phase was used on 85 subjects for determination of diastolic blood pressure.
• This device has not been validated for use in pregnancy.
Model
BP761CAN
HEM-7320T-CA
Display
LCD digital display
Measurement range
Pressure: 0 to 299 mmHg
Pulse: 40 to 180 beats / min.
Accuracy
Pressure:
±
3 mmHg or 2% of reading
Pulse:
±
5% of display reading
Inflation
Fuzzy-logic controlled by electric pump
Deflation
Automatic pressure release valve
Measurement method
Oscillometric method
IP classification
IP 20
Power source
4 “AA” batteries 1.5V or optional AC adapter (INPUT AC100-240V 50/60Hz 0.12A)
Battery life
Approximately 1000 measurements (using new alkaline batteries)
Operating temperature / humidity
50°F to 104°F (10°C to 40°C) / 15 to 90% RH
Storage temperature / humidity / air pressure
-4°F to 140°F (-20°C to 60°C) / 10 to 95% RH / 700 to 1060 hPa
Weight
Monitor : Approximately 13 5/8 oz. (385 g) not including batteries
Arm cuff : Approximately 5 3/4 oz. (163 g)
Dimensions
Monitor : Approximately 4 7/8” (w)
×
3 1/2” (h)
×
6 3/8” (l)
(124 mm
×
90 mm
×
161 mm)
Arm cuff : Approximately 5 3/4”
×
21” (air tube: 29 1/2”)
(145 mm
×
532 mm (air tube: 750 mm))
Cuff circumference
9” to 17” (220 to 420 mm)
Memory
Up to 60 per user
Contents
Monitor, ComFit™ Cuff, 4 AA batteries, instruction manual, quick start guide
Applied part
Protection against electric shock
Internally powered ME equipment (When using only the batteries)
= Type BF
= Class II ME equipment (Optional AC adapter)
HEM-7320T-Z_A_M.book Page 27 Wednesday, May 7, 2014 4:38 PM