OmniCure S2000 Elite Скачать руководство пользователя страница 4

 

035-00708R

 

R1

 

 

Installing the Optical Filter 

The OmniCure S2000 Elite curing system is equipped with user-changeable optical filters cartridges. 
Seven different filter options are available to choose from: 
400-500nm, 365nm, 320-390nm, 250-450nm, 320-500nm, no filter and customer filter. 
Choose the appropriate optical filter for your curing needs: 

1.  Ensure the AC power cord is disconnected from the unit. 
2.  Remove the screws from the lamp access side panel using the  

tool (3mm allen) provided and remove the panel from the unit cover. 

3.  Carefully remove the filter cartridge from its packaging and insert into  

the slot with the filter’s model number tag pointing towards the lamp. 

4.

 

Secure the filter to the system using the provided M4 screws using  
the 3mm allen key.

 

Note: If the filter is removed and not replaced with another filter,  
it is recommended to fasten the screws back into the system  
until another filter is installed at another time. This will prevent  
the screws from being lost. 

5.  Replace the lamp housing side panel and tighten the fastening screws. 
6.  Reconnect the AC power cord and turn on the Power switch at the back of  

the unit to turn the system ON. 

Note: If the filter is properly installed, the system will automatically recognize the type. 
It will be shown in the “Information” section of the GUI. 

Warning: If the filter is not properly installed or fastened, the filter will not be 
recognized, and the lamp will not strike 

 

Regulatory Information 

Product Safety and Electromagnetic Compatibility 

The OmniCure S2000 has been tested and found to comply with product safety and electromagnetic 
compatibility requirements.  For a complete list of tests and for certification details, please contact 
your OmniCure representative or refer to the User Guide. 
 

Council Directive 2014/35/EU 

Low Voltage Directive 

Council Directive 2014/30/EU 

EMC Directive 

Council Directive 2012/19/EU 

WEEE Directive 

Council Directive 2011/65/EU as amended by (EU) 2015/863 

RoHS 

Council Directive 2014/53/EU 

Radio Equipment 

 

FCC ID: 2AXSI-SSERIES 
IC ID: 26590-SSERIES 
 

        

                               

                     

 

 

WEEE Directive       

   China RoHS        

 

CE Marking 

 

Содержание S2000 Elite

Страница 1: ...ring will light up when power is turned on to the system followed by the screen initializing a few seconds later The lamp will not strike at startup on Demo systems by default this can be changed in the Settings Default Lamp state Note If the Default Lamp state setting has been modified to ON the lamp will strike at start up Allow the lamp to warm up uninterrupted DO NOT TURN SYSTEM OFF while the ...

Страница 2: ...tion buttons to set intensity 2 Or press the intensity window on touch screen and enter desired intensity on the number pad 3 Calibration needs to be performed to enable intensity control in W cm2 and W Locking and Unlocking the System 1 On the RUN screen select the LOCK icon and enter the PIN to lock the system Default PIN is 1234 2 When the system is locked the user cannot exit the screen or cha...

Страница 3: ...rientation If you are having difficulty attaching the connector try rotating it by 180º 7 Tuck the Intelli Lamp wires into the grommet located on the divider wall between the lamp and the connector Waring Failure to do so will result in the door not fully closing and system will not power on 8 Locate the Power connector with two leads and connect it to its mate located on the bottom of the lamp ho...

Страница 4: ... This will prevent the screws from being lost 5 Replace the lamp housing side panel and tighten the fastening screws 6 Reconnect the AC power cord and turn on the Power switch at the back of the unit to turn the system ON Note If the filter is properly installed the system will automatically recognize the type It will be shown in the Information section of the GUI Warning If the filter is not prop...

Отзывы: