OmniCure S2000 Elite Скачать руководство пользователя страница 1

 

035-00708R

 

R0

 

Excelitas Canada Inc. 

2260 Argentia Road, Mississauga ON  L5N 6H7 

Tel: 1-905-821-2600 

Toll Free: 1-800-668-8752 

www.excelitas.com/omnicure

 

 

 

 

Installing the Light Guide

 

1.

  Remove the protective end caps from the light guide.

 

2.  Insert the light guide into the light guide port on the front panel until  

it sets with a positive click. 
 

Note: When power to the unit is turned on and system is  
initialized, the light ring will change colors indicating the following states:

 

Red:

 Light Guide not detected 

Blue

: Light Guide inserted 

– System warming up

 

Magenta:

 Light Guide inserted 

– Closed-Loop Feedback Inactive 

Yellow:

 Light Guide inserted 

– Closed-Loop Feedback Active, Calibration close to expiration 

Green:

 Light Guide inserted 

– Closed-Loop Feedback Active

 

 

 

Start up 

1.  With the OmniCure

®

 S2000 Elite system not connected to power, install the lamp module (See Page 3). 

2.  Install the Optical band pass filter (See Page 4). 
3.  Plug the unit into a properly grounded AC outlet. 
4.  Turn on the Power switch at the back of the unit. 
5.  The light ring will light up when power is turned on to the system, followed by the screen initializing a few 

seconds later. 

The lamp will not strike at startup on Demo systems by default (this can be changed in 

the Settings 

– Default Lamp state).

 

 

Note: If the Default Lamp state setting has been modified to ON, the lamp will strike at start-up. 

Allow 

the lamp to warm-up uninterrupted, DO NOT TURN SYSTEM OFF while the lamp is warming up!

 

 

 

Using the OmniCure S2000 Elite 

1.

  Following the initialization period, the OmniCure S2000 Elite will display 

 

3 main options on the Home screen: Information, Run and Settings.

 

2.  Information and Settings can be accessed while the lamp is turned off. 

3.

  To start running exposures, press the Lamp icon to strike the 

 

lamp. Allow the lamp to warm-up (Excelitas recommends 20 minutes of  
proper warm-up to ensure stable optical output). During the warm-up  
period, the screen will indicate the minimum warm-up count-down.

 

Warning: UV light and elevated levels of visible light will be emitted from the end of the light guide.

 

 

S2000 Elite 

– 

Quick Start Guide 

Warning 

– Refer to Safety Precautions Booklet 035-00636R prior to use 

User Guide 035-00707R

https://www.excelitas.com/product/omnicure-s2000-Elite-spot-uv-curing-system

 

 

Содержание S2000 Elite

Страница 1: ...ring will light up when power is turned on to the system followed by the screen initializing a few seconds later The lamp will not strike at startup on Demo systems by default this can be changed in the Settings Default Lamp state Note If the Default Lamp state setting has been modified to ON the lamp will strike at start up Allow the lamp to warm up uninterrupted DO NOT TURN SYSTEM OFF while the ...

Страница 2: ...tion buttons to set intensity 2 Or press the intensity window on touch screen and enter desired intensity on the number pad 3 Calibration needs to be performed to enable intensity control in W cm2 and W Locking and Unlocking the System 1 On the RUN screen select the LOCK icon and enter the PIN to lock the system Default PIN is 1234 2 When the system is locked the user cannot exit the screen or cha...

Страница 3: ...rientation If you are having difficulty attaching the connector try rotating it by 180º 7 Tuck the Intelli Lamp wires into the grommet located on the divider wall between the lamp and the connector Waring Failure to do so will result in the door not fully closing and system will not power on 8 Locate the Power connector with two leads and connect it to its mate located on the bottom of the lamp ho...

Страница 4: ... This will prevent the screws from being lost 5 Replace the lamp housing side panel and tighten the fastening screws 6 Reconnect the AC power cord and turn on the Power switch at the back of the unit to turn the system ON Note If the filter is properly installed the system will automatically recognize the type It will be shown in the Information section of the GUI Warning If the filter is not prop...

Отзывы: