SECTION 4. SONIC RUPTOR 400 OPERATION 

4.4 TIMER SETTINGS 

The Timer is adjustable from 1 to 15 minutes. The timer is turned off when the knob is in the 
fully counter-clockwise position.  
 
When the Timer control knob is set to a desired number of minutes, the Timer LED will light 
to indicate that the Timer is on.  

4.5 PULSER SETTINGS 

The Pulser knob allows for two different power output modes:  



Constant mode  



Pulse mode  

The pulse mode applies ultrasonic energy to the tip at a rate of one pulse per second. The 
pulse duration can be adjusted from 10% to 90%, enabling a solution to be processed at full 
power range while limiting temperature. The pulse mode is especially valuable for processing 
heat sensitive samples.   
 

Example:  

Time = 5 minutes 
Pulse = 10% 
The unit will pulse every 0.5 seconds for 5 minutes.  
 
The constant mode is operative when the knob is in the fully counter clockwise position. 

4.6 START & RESET 

 
Start Button: Pressing “Start” begins the ultrasonic processing at the preset power lever, puls-
er, and time.  
 
Reset Button: Pressing “Reset” interrupts the processing. When the instrument is turned on 
but is not engaged in processing, it is in the idle state.  

Содержание Sonic Ruptor 400

Страница 1: ...ULTRASONIC HOMOGENIZER USER MANUAL OMNI SONIC RUPTOR 400 Rev A...

Страница 2: ...Ruptor 400 For Use Section 4 Sonic Ruptor 400 Operation 4 1 Control Panel 4 2 Start Up of the Equipment 4 3 Power Output 4 4 Timer Settings 4 5 Pulser Settings 4 6 Start Reset Section 5 Operation 5 1...

Страница 3: ...For the balance of the period parts will be provided but service will be charged at established labor rates This warranty does not include normal wear from use it does not apply to any instrument or...

Страница 4: ...the instrument from the main power supply remove the power cord from the AC power input on the rear panel and clean the equipment and the surrounding area DO NOT reconnect the equipment until it has...

Страница 5: ...1 Place the shipping box on a level surface 2 Remove the foam from around the control unit and accessories 3 Place the Sonic Ruptor on a clean horizontal and stable surface 4 Do not block the cooling...

Страница 6: ...N 5 32 Stepped Micro Processing Tip Very High Intensity Diameter 5 32 3 8mm Length 10 1 25 6cm 250 L 10mL OR T 156 3 8 Intermediate Processing Tip High Intensity Diameter 3 8 9 5mm Length 8 6 21 8cm...

Страница 7: ...d up to 8 micro tubes PRODUCT DESCRIPTION ORDER 250mL Cup Tip OR C 250 Microtube Tray for OR C 250 OR C 250 T PRODUCT DESCRIPTION ORDER Continuous Flow Chamber OR F 01 Continuous Flow Chamber The Cont...

Страница 8: ...apors and water CAUTION Always place a washer between the Tip and the BNC Transducer Attaching the Processing Tip Sold Separately Remove the protective plastic caps from the Processing Tip and the Tra...

Страница 9: ...power entry module located on the rear panel of the control unit Plug the power cord into a grounded outlet of 115 volts 60 cycles or 230 volts 50 cycles AC Check the Serial Plate and power entry modu...

Страница 10: ...t of ultrasonic power being generated by the Processing Tip The power output is adjusted by turning the Power knob to the right or left The ten segment LED Power Output Meter reflects the level of the...

Страница 11: ...ate of one pulse per second The pulse duration can be adjusted from 10 to 90 enabling a solution to be processed at full power range while limiting temperature The pulse mode is especially valuable fo...

Страница 12: ...vertor If the Timer is used the processing will automatically stop when the Timer is through Push the Reset button to terminate the processing at any time Transducer The Transducer assembly may be han...

Страница 13: ...ove the threshold of cavitation audible sounds are emitted from the vessel containing the liquid These sounds can be irritating to the operator and care should be taken to avoid prolonged exposure For...

Страница 14: ...the fan ALWAYS disconnect the power cord before cleaning DO NOT use any type of scrapers DO NOT use caustic solvents or acetone Harsh solvents can damage the finish of the housing Example of Decontami...

Страница 15: ...certificate and attach to the instrument or parts being returned If no Decontamination Certificate is attached and a potential radioactive or biological hazard is detected or suspected by Omni Interna...

Страница 16: ...cer 3 Replace washer The probe is not properly secured 1 Turn off the unit Use the wrenches to tighten the processing tip on the transducer Burning smell from transducer 1 Turn off the control unit 2...

Страница 17: ...sel The processing tip is not immersed in enough fluid 1 Turn off the unit 2 Check the depth of the tip in the sample liquid 3 Adjust the probe if necessary The processing tip is fractured or damaged...

Страница 18: ...earch Please visit the web address below to register your new Sonic Ruptor You will also be asked to provide us with information about your research In return we ll send you a Starbucks gift card Regi...

Страница 19: ..._________________________________________ ___________________________________________________________ ___________________________________________________________ ______________________________________...

Страница 20: ...20 935 C Cobb Place Blvd Toll Free 1 800 776 4431 Kennesaw GA 30144 Phone 770 421 0058 Patents Pending P N 03 254 www omni inc com...

Отзывы: