SECTION 2. OMNI TH, TISSUE HOMOGENIZER (cont.)
2.3 STAND-MOUNTING THE OMNI TH (Optional)
To mount the Omni TH, Tissue Homogenizer to a stand assembly, refer to
the figure and follow these steps:
1) Secure the 24” post by twisting it into
the base plate until securely tightened.
2) Loosen the locking knob and slide the
included post clamp assembly down
over the end of the post until the
clamp is at the desired height and lock
in place.
3) Slide the TH motor into the opening
on the end of the post clamp assembly
and push it down firmly. The unit will
lock in place.
4) To remove the TH, press the black
button on the back of the opening of
the post clamp assembly and lift the
motor from the cradle.
Motor Brushes
Motor Drive
Post Clamp
Assembly
Locking Knob
24” Post
Base Plate
CAUTION: When using glass-filled Teflon lower bearings, do not operate the
unit for extended periods of time without immersing the bottom of the
generator probe in liquid or the sample being processed in order to avoid
premature failure of the lower bearing.
2.1 INSTALLING ROTOR-STATOR GENERATOR PROBES
Stainless steel rotor-stator generator probes are easily installed to the Omni
TH motor by means of a quarter-turn bayonet mount. Simply push the
generator probe into the motor housing as far as possible, turn clockwise, and
release. Turn the thumb screw clockwise to tighten. Remove the blue
protective cap from the tip of the generator probe and the Omni TH is
ready to operate.
WARNING: The tip of the saw tooth generator probe is sharp.
Thumb Screw