![Omni International TIP HOMOGENIZING KIT Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/omni-international/tip-homogenizing-kit/tip-homogenizing-kit_user-manual_4004259019.webp)
19
SECTION 7. SERVICE
SECTION 8. TROUBLESHOOTING
Do not attempt to service the Omni TH in a manner other than those
discussed in this manual. For any issue that is unsuccessfully corrected using
this guide, please contact your authorized dealer or call Omni International
technical assistance at 1-800-776-4431 or 1-770-421-0058.
PROBLEM
CORRECTIVE ACTION
The Omni TH is plugged in and turned on
but is not functioning.
-Check power cord connectors
-Check wall socket for power to the
outlet
Motor is turned ON and makes a makes a
“buzzing” sound, but is not working.
-Check that the brushes are not worn
and are correctly installed. Replace if
necessary. (see Section 3.2)
Motor unit operating speed declines, stalls
intermittently, or stops completely
-Check that the brushes are not worn
and are correctly installed. Replace if
necessary. (see Section 3.2)
Teflon bearing wears quickly
-Fluid level may be too low in the tube.
-Immerse the probe deeper into the fluid.
Excessive splashing in sample tube
-Fluid level too low for tube size
7.1 A
SSISTANCE
Should this product ever require service, please contact Omni International at
1–800–776-4431.
7.2 D
ECONTAMINATION
Should an instrument or component that has been used with radioactive or
pathogenic material require factory or field service, comply with the following
procedure to ensure the safety of service personnel:
Clean the parts to be serviced of all encrusted material and decontaminate
them. There must be no radioactivity detectable by survey equipment.
Obtain a Decontamination Certificate from Omni International. Complete
the certificate and attach to the instrument or parts being returned.
If no Decontamination Certificate is attached, and a potential radioactive or
biological hazard is detected or suspected by Omni International, the equipment
will not be serviced until proper decontamination and certification is complete.
The sender will be contacted for instructions as to the disposition of the
equipment. Disposition costs will be borne by the sender.
WARNING:
It is a violation of federal law to transport biologically
hazardous or radioactive materials without proper packaging, labeling, and
appropriate warnings.