Error Code
Definition
Cause
Possible Solution
Error 1
Display screen
shows
incomplete or
unusual characters
CPU needs to be
reset
Turn off the unit. Wait 30 seconds.
Turn the unit back on.
Error 2
No display on the
screen
No Power
Check main voltage.
Check that the voltage of the unit (100V-240V)
matches what is being delivered by the main
power supply.
Make sure the unit is plugged in properly.
Check and replace fuses if blown.
Faulty display screen
Turn off the unit.
Contact manufacturer for assistance.
Error 3
Unit powers on but
motor does not turn
Faulty motor
Contact manufacturer for assistance.
Error 4
Close Door
Lid is not seated
properly.
Check Lid latch for obstruction
Press lid and ensure the handle is closed
Detection system
is faulty.
Troubleshooting
Error Codes
WARNING: Any service must be performed by a qualified service technician. This can be either
an OMNI Technician or an end user resource.
All parts need to be sourced from the OMNI International service department
ADVERTISEMENT : Tout entretien doit être effectué par un technicien d’entretien qualifié. Il peut
s’agir d’un technicien OMNI ou d’une ressource utilisateur final.
Toutes les pièces doivent provenir du service après-vente d’OMNI International.
Содержание Bead Ruptor 4
Страница 1: ...User Manual Bead RuptorTM 4 Bead Mill Homogenizer...
Страница 2: ......
Страница 14: ...Notes...
Страница 15: ...Notes...
Страница 16: ...PN 03 225 01 Rev N 935 Cobb Place Blvd Kennesaw GA 30144 800 776 4431 770 421 0058 www omni inc com...