Wiring
You can connect
VBO3
to
CMC 430
test sets as well as to other OMICRON CMC test sets
(
CMC 256plus
,
CMC 310
,
CMC 353
, and
CMC 356
). Note, however, that the wiring and the
configuration in the
Test Universe
software are different depending on the respective use case. For
more information on how to configure the control software, see page
WARNING
Avoid serious injury or death caused by electric shock.
► The isolation of
VBO3
is designed as a 600 V isolation. Therefore, only operate
VBO3
with
primary voltages
up to max. 300 V. Otherwise, output voltages higher
than 600 V might occur due to the voltage ratio of about 1 : 2.16.
► Always use the connection cables with 4 mm safety banana plugs (that are
included in the delivery) to connect
VBO3
to a CMC test set.
► If you need to use other cables and/or connectors, always refer to the safety
instructions and user documentation of the respective CMC test set in use before
proceeding.
Wiring
OMICRON
11