Technical Data
61
Table 6-17:
Data for potential-sensing
operation
Table 6-18:
Data for potential-free
operation
Data for Potential-Sensing Operation
Threshold voltage data per input
range
Setting range
Resolution
Range I
Range II
0...20V
>20...300V
50mV
500mV
Max. input voltage
CAT III/ / 300 V
rms
CAT IV / 150 V
rms
Threshold voltage accuracy
1
1
Applies to positive voltage signal edge; value shown in % of reading (rd.) + % of upper range
value (rg.)
5% of rd. + 0.5% of rg.
Threshold voltage hysteresis
Range I:
typ. 60 mV
Range II: typ. 900 mV
Input impedance
2
Threshold 0...20V
Threshold 20...300V
2
210 k
Ω
135 k
Ω
Data for Potential-Free Operation
1
1
Trigger criteria
Logical 0: R > 100 k
Ω
Logical 1: R < 10 k
Ω
Input impedance
216 k
Ω
Содержание CMC 356
Страница 1: ...CMC 356 Reference Manual ...
Страница 40: ...CMC 356 Reference Manual 40 ...
Страница 104: ...CMC 356 Reference Manual 104 ...
Страница 108: ...Open Source License Information 108 ...
Страница 110: ...Support 110 ...