OMICRON Lab
5
Safety
2.3
Safety instructions
Before operating the
B-AMP 12
, read the safety instructions in this document carefully. If you do not
understand some safety rules, contact OMICRON Lab before proceeding.
Before operating always make sure that neither the
B-AMP 12
nor its accessories are damaged.
Test setups that include the
B-AMP 12
must comply with the internal safety instructions and all
additional safety relevant documents.
For your safety
The
B-AMP 12
may only be operated by trained personnel. Any maloperation can result in damage to
property or persons.
Rules for use
► B-AMP 12 is a SELV device (Safety Extra Low Voltage in accordance to IEC 60950), also known
as protection class III equipment.
► Do not apply hazardous voltage levels >50 VDC or >25 VAC to any input or output of
B-AMP 12.
► Be aware that no indication on the B-AMP 12 shows that the output is active. This could be
especially critical if additional amplifiers are connected to the B-AMP 12.
► Testing with the
B-AMP 12
shall only be carried out by authorized and qualified personnel. Before
starting to work, clearly establish the responsibilities of all personnel involved.
► Do not operate the B-AMP 12 when explosive gas or vapours are present.
2.4
Cleaning
To clean the B-AMP 12, use a cloth dampened with isopropanol alcohol. Prior to cleaning, always ensure
that all connections to the B-AMP 12 and DUT are disconnected.