No lights on the control panel.
1. Is the Power push button pushed in?
2. Is there power to the facility outlet?
3. Turn off Power and unplug the machine.
4. Remove upper panel located on the right end of the machine.
5. Is the circuit breaker energized (UP-Red) or (DOWN-Green)?
6. If the circuit breaker is tripped, re-set the circuit breaker.
7. Plug in machine.
8. Follow the steps above in
“Isolate the exact issue”.
9. Machines manufactured after January 2015 have panel fuses and may be
burnt. If either of the fuse holder red lights are lit, the 5 amp fuse(s) must
be replaced.
Heat is causing the circuit breaker to trip.
1. Likely cause is a shorted heat blanket.
2. Remove the front (if applicable) and back lower access panels.
3. Inspect the heat blanket(s)
– they should be an orange/ pinkish color.
4. If the heat blankets are black or charcoal white, they must be replaced.
5. Call Omegasonics to order new heat blanket(s).
Pump is causing the circuit breaker to trip.
1. Likely cause is a faulty or shorted motor.
2. Call Omegasonics to order a new pump.