3)Pressing “Offset” button (3-8, Fig. 1)
for at least two seconds then release, the
display will show :
SEt
oFS
Pt 1
Pt 2
4)If you wish to make the offset adjustment for Pt 1, please
insert the probe to PT1 input socket. Press the “L” button
(3-5, Fig. 1) once, the display will show example as following.
If you wish to make the offset adjustment for Pt 2, please
insert the probe to PT2 input socket. Press the “R” button
( 3-6, Fig. 1 ) once, the display will show example as
following.
SEt
oFS
23.7
23.7
* If no probe is inserted
into PT1, PT2 input
measuring
adjust
socket, the meter
value
value
will show “Err”
* Use the “
▲
” button (3-5, Fig. 1) “
▼
” button (3-6,
Fig. 1) to adjust the desired value on right bottom
display.
* Pressing “Enter” button (3-4, Fig. 1) once, the
adjustment value will be saved into the memory
then return to normal measuring screen to finish
the offset adjustment procedures.
* The above offset adjustment for Pt 100 ohm is
valid for PT1, PT2 individually.
23