![Omega PHCN-962 Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/omega/phcn-962/phcn-962_user-manual_4183689036.webp)
36
2) RE (read the appointment data)
The PC send : 40,30,31,52,45,30,30,adrH,adrL,lthH,lthL,CRH,CRL,0D (total
14 bytes)
40 : the start byte
30 31 : the ID of instrument (the hexadecimal: 0
~
3FH, ASCII
:
30 30
~
33 46
);
52 45 : the instruction R, E;
30 30 :reserve
adrH adrL : the address of the start parameter byte(the hexadecimal: 0
~
1BH
;
ASCII
:
30 30
~
31 42
);
1theH 1theL : the count of the parameters (the hexadecimal: 1
~
1CH
;
ASCII
:
30 31
~
31 43
);
CRCH CRCL : CRC
0D : the end byte
The adr is the start address(00
~
1BH
)
,1th is the count of the byte. The
table is the relation of the adr and parameter.
adr
00
01
02
03
04
05
06
07
08
09
0A
0B
0C
0D
para
TST1
TST1
TST2
TST2
SP1U
SP1U
SP2U
SP2U
CURL
CURL
CURH
CURH
POFS
POFS
adr
0E
0F
10
11
12
13
14
15
16
17
18
19
1A
1B
para
HOR
HOR
AAA
FUNC
R3OP
TOFS
SEC
SP1D
SP1T
SP2D
SP2T
NB
BT
CONF
If adr>1BH or adr+1th>1BH, the return the error code: 40,30,31,52,45,2A,2A,CRH,CRL,0D
The 2A 2A is the error sign
If the address range is correct then the instrument will send back data :
40,30,31,52,45,D1,D2,D3,
………
Dn,CRH,CRL,0D
40:the start byte
30 31 : the ID
52 45 : the fix data
D1
~
Dn
:
parameters from instrument to PC
CRCH
,
CRCL:CRC
0D : the end byte
3) RR (read all of parameters) the PC send : 40,30,31,52,52,CRH,CRL,0D(8 bytes)
40: the start byte
30 31 : the ID
52 52 : the instruction of R, R
CRCH,CRCL :CRC
0D : the end byte