Press
twice to quit one-point calibration and return to measuring status. Press
to continue
next point calibration.
Take electrode out of the first standard solution, clean it and put it into the second standard
solution.
5.
If select one point calibration, instrument will show the slope, but zero point adopts new calibration
value while the slope remains the value of last calibration. If select two point calibration, the lower
display will show the next value of standard buffer solution. Use
or
key
to selects the
second standard buffer solution from the pre-set standard buffer solution. Take electrode out of
the first standard solution, clean it and put it into the second standard solution. Press
key
,
BUFF
will blink. Instrument will modify the indication to the value of standard solution.
6.
After finish the second calibration of standard buffer solution, instrument will automatically show
SLOP on LCD main display zone. The right lower display shows the slope of electrode.
Parameter of zero point and slope will be renewed after each calibration.