39
6.2
PARAMETER SETTING AND FACTORY PRELIMINARY VALUE
NO.
Indication
Parameter
Name
Symbol
Contents
Remark
Valid range
Factor
y value
01
LOCK
LOC
Password for entering function
0
~
200
0
02
ATC
AtC
Auto temperature compensation
ON/OFF
OFF
03
TSET1
tSt1
Temperature set up of manual
-10.0
~
100.0
°C
25.0
04
TSET2
tSt2
Calibration temperature set up
Only valid for manual
0.0
~
60.0
°C
25.0
05
P1
TC
TOFS
tOFS
Temperature measurement offset
Only valid for manual
-10.0
~
10.0
°C
0.0
POFS
POFS
PH value measurement offset
Only valid for PH mode
-2.00
~
2.00PH
0.00
06
P2
OFS
VOFS
VOFS
mV value measurement offset
Only valid for ORP mode
-200
~
200mV
0
07
CTYP
CtyP
Type of current output
0/4
~
20mA
4
08
CURL
CUrL
Low limit of transmitting output
-1999
~
1999
0
09
P3
CUr
CURH
CUrH
High limit of transmitting output
PH mode: 0.01PH unit
ORP mode: 1mV unit
-1999
~
1999
1400
10
P4Cntr
CNTR
Cntr
Control way of relay 1, 2
PLC/ Lit
Lit
11
SP1
SP1
Way set up of relay 1
HI / LO
LO
12
SP1U
SP1U
Setting value of relay 1
-1999
~
1999
400
13
SP1D
SP1d
Relay 1 hysteresis / proportion
PH mode: 0.01PH unit
ORP mode: 1mV unit
0
~
200
(
2.00
)
10
14
P5
SP1
SP1T
SP1t
Relay 1 circle (second)
Only valid for PLC
0
~
200 seconds
20
15
SP2
SP2
Way set up of relay 2
HI / LO
HI
16
SP2U
SP2U
Setting value of relay 2
-1999
~
1999
1000
17
SP2D
SP2d
Relay 2hysteresis / proportion
PH mode: 0.01PH unit
ORP mode: 1mV unit
0
~
200
(
2.00
)
10
18
P6
SP2
SP2T
SP2t
Relay 1 circle (second)
Only valid for PLC
0
~
200 seconds
20
19
R3OP
r3OP
Working mode of relay 3
OFF/SP1/SP2/
ALL/CAL/CLE
CAL
20
INT
HOr
Interval (hour)
valid for calibration and
cleaning mode
0-999
100
21
P7
rL3
DUR
SEC
Operating time (second)
valid for cleaning mode
0-200
30
22
FUNC
FUnC
PH / ORP/mV select
PH/ORP/mV
PH
23
SENS
SenS
Antimony / glass electrode select
Anti/GLAS
GLAS
24
P8
CONF
BUFF
buFF
USA/nST standard select
Only valid for PH mode
NST/USA
USA
25
NB
nb
Set ID number for 485
0-63
1
26
P9
ID
BT
bt
Communication rate
0-7
5
27
P10
DEF
DEF
dEF
Reverting to factory default
setting
Reset to factory default
parameter.
YES/NO
NO
Содержание PHCN-961